Instructions
RIGOL  Chapter 2 Command Systems for GPSA and RTSA Modes 
2-4  RSA3000E Programming Guide 
:CALCulate:BANDwidth|BWIDth:NDB 
Syntax 
:CALCulate:BANDwidth|BWIDth:NDB <rel_ampl>   
:CALCulate:BANDwidth|BWIDth:NDB? 
Description 
Sets the N value in N dB BW measurement.  
Queries the N value in N dB BW measurement.
Parameter 
Name 
Type 
Range 
Default 
<rel_ampl> 
Consecutive Real Number 
-140 dB to -0.01 dB 
-3.01 dB 
Remarks 
This command is invalid for the PvT measurement mode in RTSA. 
Return Format 
The query returns the N value in scientific notation. 
Example 
The following command sets N to -4. 
:CALCulate:BANDwidth:NDB -4 
:CALCulate:BWIDth:NDB -4 
The following query returns -4.000000e+00. 
:CALCulate:BANDwidth:NDB?
:CALCulate:BWIDth:NDB? 
:CALCulate:BANDwidth|BWIDth:RESult? 
Syntax 
:CALCulate:BANDwidth|BWIDth:RESult?
Description 
Queries the measurement results of the N dB band, and its unit is Hz.
Remarks 
This command is invalid for the PvT measurement mode in RTSA. 
Return Format 
The query returns the bandwidth value in scientific notation (with Hz as the unit).
If the two points which are located at both sides of the current marker with N dB fall or rise in its peak 
amplitude are not found, the query returns -1.000000000e+02. 