Instructions
Chapter 2 Command Systems for GPSA and RTSA Modes  RIGOL 
RSA3000E Programming Guide  2-209 
:TRIGger:SEQuence:EXTernal1:DELay:STATe ON or :TRIGger:SEQuence:EXTernal1:DELay:STATe 1 
The following query returns 1. 
:TRIGger:SEQuence:EXTernal1:DELay:STATe? 
:TRIGger[:SEQuence]:EXTernal<n>:SLOPe 
Syntax 
:TRIGger:SEQuence:EXTernal<n>:SLOPe POSitive|NEGative 
:TRIGger:SEQuence:EXTernal<n>:SLOPe?
Description
Sets the trigger edge for the external trigger. 
Queries the trigger edge for the external trigger. 
Parameter
Name 
Type 
Range 
Default 
<n> 
Discrete 
1|2 
—— 
—— 
Keyword 
POSitive|NEGative 
POSitive 
Remarks 
When the parameter n is set to 1, it indicates External Trigger 1; when set to 2, it indicates External Trigger 
2. 
POSitive: indicates the rising edge.
NEGative: indicates the falling edge. 
Return Format
The query returns POS or NEG. 
Example
The following command sets the trigger edge of External Trigger 1 to Positive.   
:TRIGger:SEQuence:EXTernal1:SLOPe POSitive 
The following query returns POS. 
:TRIGger:SEQuence:EXTernal1:SLOPe? 
:TRIGger[:SEQuence]:FMT:APTRigger 
Syntax 
:TRIGger[:SEQuence]:FMT:APTRigger <number> 
:TRIGger[:SEQuence]:FMT:APTRigger? 
Description
Sets the number of times for acquisition after each effective trigger is completed. 
Queries the number of times for acquisition after each effective trigger is completed. 
Parameter
Name 
Type 
Range 
Default 
<number> 
Integer 
Refer to "Remarks" 
1 
Remarks 
The minimum value of the parameter <number> is 1. When the trigger criteria of FMT is "Inside" or 