Instructions
Chapter 2 Command Systems for GPSA and RTSA Modes  RIGOL 
RSA3000E Programming Guide  2-181 
:STATus:OPERation:CONDition? 
Syntax
:STATus:OPERation:CONDition? 
Description
Queries the condition register of the Operation Status Register. 
Return Format
The query returns the condition register in integer. For example, 24. 
:STATus:OPERation:ENABle 
Syntax 
:STATus:OPERation:ENABle <integer> 
:STATus:OPERation:ENABle? 
Description
Sets the enable register of the Operation Status Register. 
Queries the enable register of the Operation Status Register.
Parameter
Name 
Type 
Range 
Default 
<integer>  Integer 
Refer to "Remarks" 
0 
Remarks
The following table lists the bit definitions for the operation status register. The bit 0 to bit 2, bit 5 to bit 7, 
bit 13 and bit 14 are reserved; they can be set but will not affect the instrument. The bit 15 and bit 12 to bit 
8 are not used and are always treated as 0; therefore, the range of <integer> are the decimal numbers 
corresponding to the binary numbers ranging from 0000000000000000 (0 in decimal) to 
1111111111111111 (32767 in decimal) and of which the bit 15 and bit 12 to bit 8 are 0.
Bit 
Value 
Definition 
0 
1 
Reserved 
1 
2 
Reserved 
2 
4 
Reserved 
3 
8 
SWEeping 
4 
16 
MEASuring 
5 
32 
Reserved 
6 
64 
Reserved 
7 
128 
Reserved 
8 
0 
Not Used 
9 
0 
Not Used 
10 
0 
Not Used 
11 
0 
Not Used 
12 
0 
Not Used 
13 
8192 
Reserved 
14 
16384 
Reserved 
15 
0 
Not Used 
Return Format
The query returns the enable register of the Operation Status Register in integer. 