Instructions
RIGOL  Chapter 2 Command Systems for GPSA and RTSA Modes 
2-156  RSA3000E Programming Guide 
[:SENSe]:MCHPower:FREQuency:SPAN 
Syntax 
[:SENSe]:MCHPower:FREQuency:SPAN <freq>   
[:SENSe]:MCHPower:FREQuency:SPAN?
Description 
Sets the channel span of the multi-channel power measurement.  
Queries the channel span of the multi-channel power measurement.
Parameter
Name 
Type 
Range 
Default 
<freq> 
Consecutive Real Number 
10 Hz to 2*Min[(CF - start frequency), 
(stop frequency - CF)] 
3 GHz 
Remarks
This command is only valid when the multi-channel power measurement is enabled in GPSA mode. 
Return Format
The query returns the channel span in scientific notation. Its unit is Hz. 
Example 
The following command sets the channel span to 1 MHz. 
:SENSe:MCHPower:FREQuency:SPAN 1000000 
The following query returns 1.000000000e+06.  
:SENSe:MCHPower:FREQuency:SPAN? 
[:SENSe]:MCHPower:SHEet 
Syntax 
[:SENSe]:MCHPower:SHEet OFF|ON|0|1 
[:SENSe]:MCHPower:SHEet? 
Description
Opens or closes the channel editing table. 
Queries the status of the channel editing table. 
Parameter
Name 
Type 
Range 
Default 
-- 
Bool 
OFF|ON|0|1 
OFF|0 
Remarks 
This command is only valid when the multi-channel power measurement is enabled in GPSA mode.   
Return Format
The query returns 0 or 1. 
Example
The following command enables the channel editing table.  
:SENSe:MCHPower:SHEet ON or :SENSe:MCHPower:SHEet 1 
The following query returns 1.  
:SENSe:MCHPower:SHEet? 