Trace Display 
Enables or disables the trace display. 
Remarks: 
  The trace state and the parameter settings for the corresponding state are as 
follows: 
  Active: Trace Update and Trace Display are "On". 
  View: Trace Update is "Off", and Trace Display is "On". 
  Clear: Trace Update and Trace Display are "Off". 
  Back-end: Trace Update is "On", and Trace Display is "Off". 
  In most cases, the inactive trace data keep unchanged. However, the data may 
be changed in the following conditions: 
  Load the stored data to trace. 
  Trace data are cleared by the trace clear function. 