Instructions
RIGOL  Chapter 3 Functions of the Front Panel of RTSA 
3-34  RSA3000E User Guide 
Input/Output 
Sets the input/output interface of the instrument. 
Input Impedance 
Sets the input impedance for voltage-to-power conversions (refer to equation (2-4). 
The default input impedance is 50 Ω. If the output impedance of the system under 
measurement is 75 Ω, you should use a 75 Ω to 50 Ω adapter (option) supplied by 
RIGOL to connect the analyzer with the system under test, and then set the input 
impedance to 75 Ω. 
Ext Gain 
Compensates for gain or loss in the measurement system outside the instrument. 
Remarks: 
  The value does not affect the trace position, but will modify the parameters such 
as reference level readout and the marker amplitude readout. 
  You can use the numeric keys, the knob, or arrow keys to modify this parameter; 
you can also use the touch screen to modify the parameter. For details, refer to 
descriptions in "Parameter Setting". 
Table 3-20 External Gain 
Parameter 
Remarks 
Default 
0 dB 
Range 
-120 dB to 120 dB  
Unit 
dB 
Knob Step 
1 dB 
Left/Right Arrow Key Step 
Up/Down Arrow Key Step 
5 dB 
Ext Trigger2 
Sets the external trigger 2 [TRIGGER IN/OUT] to be the "input" or "output" 
interface. 