User manual
RIGOL Chapter 2 Command System
2-186 RSA3000 Programming Guide
:SYSTem:PRESet:USER:SAVE
Syntax
:SYSTem:PRESet:USER:SAVE
Description
Saves the current user setting.
Remarks
If the current preset type is "FACTory", save the current user setting to User1. If the current preset type is
"User1 through User6", save the current user setting to the specified user.
:SYSTem:SCPI:DISPlay
Syntax
:SYSTem:SCPI:DISPlay OFF|ON|0|1
:SYSTem:SCPI:DISPlay?
Description
Enables or disables the SCPI display.
Queries the status for the SCPI display.
Parameter
Name
Type
Range
Default
——
Bool
OFF|ON|0|1
OFF|0
Remarks
When you use the SCPI commands to control the instrument remotely, turning on the SCPI display will go to
the specified menu of the corresponding command; turning off the SCPI Display will not go to the specified
menu.
Return Format
The query returns 0 or 1.
Example
The following command enables the SCPI display.
:SYSTem:SCPI:DISPlay 1 or :SYSTem:SCPI:DISPlay ON
The following query returns 1.
:SYSTem:SCPI:DISPlay?
:SYSTem:SHOW
Syntax
:SYSTem:SHOW OFF|SYSTem|OPTion|LICense
:SYSTem:SHOW?
Description
Displays the system-related information.
Queries the system-related information.