User manual
RIGOL Chapter 2 Command System
2-154 RSA3000 Programming Guide
[:SENSe]:SIGCapture:2FSK:AMPUp
Syntax
[:SENSe]:SIGCapture:2FSK:AMPUp <ample>
[:SENSe]:SIGCapture:2FSK:AMPUp?
Description
Sets the upper limit of the amplitude of 2FSK signal.
Queries the upper limit of the amplitude of 2FSK signal.
Parameter
Name
Type
Range
Default
<ample>
Consecutive Real Number
-100 dBm to 320 dBm
-10 dBm
Remarks
This command is only available for RTSA mode and valid when the SSC function is enabled.
Return Format
The query returns the amplitude upper limit in scientific notation.
Example
The following command sets the amplitude upper limit to -20 dBm.
:SENSe:SIGCapture:2FSK:AMPUp -20
The following query returns -2.000000000e+01.
:SENSe:SIGCapture:2FSK:AMPUp?
[:SENSe]:SIGCapture:2FSK:MARK1:FREQ
Syntax
[:SENSe]:SIGCapture:2FSK:MARK1:FREQ <freq>
[:SENSe]:SIGCapture:2FSK:MARK1:FREQ?
Description
Sets the frequency value at Marker 1.
Queries the frequency value at Marker 1.
Parameter
Name
Type
Range
Default
<freq> Consecutive Real Number (start frequency) to (stop frequency)
Start
Frequency
Remarks
This command is only available for RTSA mode and valid when the SSC function is enabled.
Return Format
The query returns the frequency value of Marker 1 in scientific notation. The unit is Hz.
Example
The following command sets the frequency value at Marker 1 to 1 MHz.
:SENSe:SIGCapture:2FSK:MARK1:FREQ 1000000
The following query returns 1.000000000e+06.
:SENSe:SIGCapture:2FSK:MARK1:FREQ?