User manual
Chapter 2 Command System RIGOL
RSA3000 Programming Guide 2-147
Return Format
The query returns the average count in integer.
Example
The following command sets the average count to 100.
:SENSe:MCHPower:AVERage:COUNt 100
The following query returns 100.
:SENSe:MCHPower:AVERage:COUNt?
[:SENSe]:MCHPower:AVERage[:STATe]
Syntax
[:SENSe]:MCHPower:AVERage[:STATe] OFF|ON|0|1
[:SENSe]:MCHPower:AVERage[:STATe]?
Description
Enables or disables the average measurement function of the multi-channel power measurement.
Queries the status of the average measurement function of the multi-channel power measurement.
Parameter
Name
Type
Range
Default
——
Bool
OFF|ON|0|1
ON|1
Remarks
This command is only valid when the multi-channel power measurement in GPSA mode is enabled.
Return Format
The query returns 0 or 1.
Example
The following command enables the average measurement function.
:SENSe:MCHPower:AVERage:STATe ON or :SENSe:MCHPower:AVERage:STATe 1
The following query returns 1.
:SENSe:MCHPower:AVERage:STATe?
[:SENSe]:MCHPower:AVERage:TCONtrol
Syntax
[:SENSe]:MCHPower:AVERage:TCONtrol EXPonential|REPeat
[:SENSe]:MCHPower:AVERage:TCONtrol?
Description
Selects the average mode of the multi-channel power measurement.
Queries the average mode of the multi-channel power measurement.
Parameter
Name
Type
Range
Default
——
Keyword
EXPonential|REPeat
EXPonential