RIGOL Chapter 2 Command System
2-146 RSA3000 Programming Guide
Example
The following command sets the number of harmonics to 5.
:SENSe:HDISt:NUMBers 5
The following query returns 5.
:SENSe:HDISt:NUMBers?
[:SENSe]:HDISt:TIME
Syntax
[:SENSe]:HDISt:TIME <time>
[:SENSe]:HDISt:TIME?
Description
Sets the sweep time of the harmonic distortion measurement.
Queries the sweep time of the harmonic distortion measurement.
Parameter
Name
Type
Range
Default
<time>
Consecutive Real Number
1 μs to 6 ks
1 ms
Remarks
This command is only valid when the harmonic distortion measurement in GPSA mode is enabled.
This setting changes the sweep time of the analyzer.
Return Format
The query returns the sweep time in scientific notation. The unit is s.
Example
The following command sets the sweep time to 100 ms.
:SENSe:HDISt:TIME 0.1
The following query returns 1.000000000e-01.
:SENSe:HDISt:TIME?
[:SENSe]:MCHPower:AVERage:COUNt
Syntax
[:SENSe]:MCHPower:AVERage:COUNt <integer>
[:SENSe]:MCHPower:AVERage:COUNt?
Description
Sets the average count of the multi-channel power measurement.
Queries the average count of the multi-channel power measurement.
Parameter
Name
Type
Range
Default
<integer>
Integer
1 to 1,000
10
Remarks
This command is only valid when the multi-channel power measurement in GPSA mode is enabled.