RIGOL Chapter 2 Command System
2-144 RSA3000 Programming Guide
Parameter
Name
Type
Range
Default
<integer>
Integer
1 to 1,000
10
Remarks
This command is only valid when the harmonic distortion measurement in GPSA mode is enabled.
Return Format
The query returns the average count in integer.
Example
The following command sets the average count to 100.
:SENSe:HDISt:AVERage:COUNt 100
The following query returns 100.
:SENSe:HDISt:AVERage:COUNt?
[:SENSe]:HDISt:AVERage[:STATe]
Syntax
[:SENSe]:HDISt:AVERage[:STATe] OFF|ON|0|1
[:SENSe]:HDISt:AVERage[:STATe]?
Description
Enables or disables the average measurement function of the harmonic distortion measurement.
Queries the status of the average measurement function of the harmonic distortion measurement.
Parameter
Name
Type
Range
Default
——
Bool
OFF|ON|0|1
OFF|0
Remarks
This command is only valid when the harmonic distortion measurement in GPSA mode is enabled.
Return Format
The query returns 0 or 1.
Example
The following command enables the average measurement function.
:SENSe:HDISt:AVERage:STATe ON or :SENSe:HDISt:AVERage:STATe 1
The following query returns 1.
:SENSe:HDISt:AVERage:STATe?