User manual
RIGOL Chapter 2 Command System
2-136 RSA3000 Programming Guide
Example
The following command sets the average mode to Repeat.
:SENSe:EBWidth:AVERage:TCONtrol REPeat
The following query returns REP.
:SENSe:EBWidth:AVERage:TCONtrol?
[:SENSe]:EBWidth:FREQuency:SPAN
Syntax
[:SENSe]:EBWidth:FREQuency:SPAN <freq>
[:SENSe]:EBWidth:FREQuency:SPAN?
Description
Sets the span of the emission bandwidth measurement.
Queries the span of the emission bandwidth measurement.
Parameter
Name
Type
Range
[1]
Default
<freq>
Consecutive Real Number
100 Hz to 4.5 GHz
2 MHz
Note
[1]
: If option RSA3000-BW1 is installed, the available range is from 10 Hz to 4.5 GHz.
Remarks
This command is only valid when the emission bandwidth measurement in GPSA mode is enabled.
This setting changes the span of the analyzer.
Return Format
The query returns the span in scientific notation. Its unit is Hz.
Example
The following command sets the span to 10 MHz.
:SENSe:EBWidth:FREQuency:SPAN 10000000
The following query returns 1.000000000e+07.
:SENSe:EBWidth:FREQuency:SPAN?
[:SENSe]:EBWidth:MAXHold:STATe
Syntax
[:SENSe]:EBWidth:MAXHold:STATe OFF|ON|0|1
[:SENSe]:EBWidth:MAXHold:STATe?
Description
Enables or disables the EBW Max Hold.
Queries the status of the EBW Max Hold.
Parameter
Name
Type
Range
Default
——
Bool
OFF|ON|0|1
OFF|0
Remarks
This command is only valid when the emission bandwidth measurement in GPSA mode is enabled.