Information
RSAxxx(N) EMI Mode – Getting Started
RIGOL Technologies CO., LTD. 09.09.2021
Figure 8: Result of measurement after finalization of test with the settings described above
Therefore you must select separately the dynamic range of the measurement meter as follow:
Please select [AMPT] and adjust attenuator and pre-amplifier / reference level according to your
need (see figure 7).
Correction:
In case you using external components like attenuators, amplifiers, cable, LISN, etc. you can add a
correction which will be take into account into the result of trace, measurement meter and signal table.
To insert correction, please select [Input Output] and [Corrections]
Start The Measurement:
After you finalized all the setup of your need, you can start the test.
Please select [Sweep] [Start]
Signal Table Settings:
Signal table displays all the peaks which are detected in the trace according to your adjusted search
criteria.
The numbers of detected peaks can be adjusted in the signal table as follow: Select [Meas Setup]
[Signal Table] and insert the settings for search criteria and sort of signals according to your
need.
You can also select one peak and [Zoom In] for deeper analysis.