RSAxxx(N) EMI Mode – Getting Started
Introduction:
In this description you will find first steps of operation how to work with EMI Mode in RIGOL’s RSA Series
1
and how to adjust different settings. The target should be, to understand the principle of this mode for
pre-compliance analysis and perform easily adequate settings from the beginning.
The principle of EMI mode is, to make a pre-compliance analysis (conducted or radiated emission) and safe
test time for that evaluation. In case a whole trace would be recorded with a valued detector (like quasi
peak) the test time would be quite long. This EMI mode can do a fast scan of the trace with e.g. [Average]
or [Peak] detector and search all peaks. With that peaks, the measurement of valued detectors (QP and/or
CISPR AV) can be done and the result is available within minutes.
But saving test time is only one part. Failure debugging and design improvement is a second step. With
the integrated measurement meter, customer can do a live debugging and can do an immediate
improvement of each unwanted single peak. Last but not least, the mode allows you to generate a test
report for your documentation. This mode contents a lot of features, for correction, pass / fail testing, and
much more.
Default Window, EMI – Mode:
Activate the EMI Mode and see the window of figure 1.
The default mode is set to CISPR filter (6 dB filter). In case that for some evaluation 3 dB filter are required,
please change this here: Please select [Mode Setup] [EMC Standard]: “None”.
1