Datasheet

RIGOL
2
Specications
Measurement Mode
All Measurement Modes
Specications are valid under the following conditions: the instrument is within the calibration period, is stored for at least two hours at 0
to 50
temperature, and is warmed up for 40 minutes. Unless otherwise noted, the specications in this manual include the measurement
uncertainty.
Typical: characteristic performance, which 80 percent of the measurement results will meet at room temperature (approximately 25
). This
data is not warranted and does not include the measurement uncertainty.
Nominal: the expected mean or average performance or a designed attribute (such as the 50 Ω connector). This data is not warranted and is
measured at room temperature (approximately 25
).
Measured: an attribute measured during the design phase which can be compared to the expected performance, such as the amplitude drift
variation with time. This data is not warranted and is measured at room temperature (approximately 25
).
NOTE:
All charts in this manual are the measurement results of multiple instruments at room temperature unless otherwise noted. The specications (except the
tracking generator specications) listed in this manual are those when the tracking generator is off.
Measurement Mode
General-Purpose Spectrum Analyzer (GPSA)
Real-time Spectrum Analyzer (RTSA)
Frequency
RSA3030 RSA3030-TG RSA3045 RSA3045-TG
Frequency Range 9 kHz to 3.0 GHz 9 kHz to 4.5 GHz
Internal Reference Frequency
Reference Frequency 10 MHz
Accuracy ±[(time since last calibration × aging rate) + temperature stability + calibration accuracy]
Initial Calibration
Accuracy
Standard <1 ppm
Option OCXO-C08 <0.1 ppm
Temperature
Stability
0
to 50
, with the reference 25
Standard <0.5 ppm
Option OCXO-C08 <0.005 ppm
Aging Rate
Standard <1 ppm/year
Option OCXO-C08 <0.03 ppm/year