Table 2-27 Tracking Generator Output Amplitude Offset 
Parameter 
Remarks 
Default 
0 dB 
Range 
-200 dB to 200 dB 
Unit 
dB 
Knob Step 
1 dB 
Left/Right Arrow Key Step 
Up/Down Arrow Key Step 
10 dB 
Normalize 
Before enabling the normalize function, connect the tracking generator output 
terminal [Gen Output 50Ω] to the RF input terminal [RF Input 50Ω]. 
Normalization can eliminate the error of tracking generator output amplitude. 
Note: Only when Output under the TG menu is "On" and the current Y-axis scale is 
log, can the Normalize menu be enabled. 
1.  Normalize 
Enables or disables the normalize function. If no saving operation is performed 
on the reference trace before enabling the normalize function, the next time you 
enable the function, the analyzer will save the reference trace automatically 
after it completes the current sweep. During saving the reference trace, a 
prompt message will be displayed. When the normalize function is enabled, the 
corresponding value of the reference trace will be subtracted from the trace data 
after every sweep. 
2.  Reference Level 
After the normalize function is enabled, you can adjust the vertical position of 
the trace on the screen by adjusting the reference level. 
  Different from the Ref Level sub-menu under AMPT, the parameter 
"Reference Level" under "Normalize" will not affect the reference level of 
the analyzer once it is modified. 
  You can use the numeric keys, the knob, and the arrow keys on the front 
panel to modify this parameter; also you can modify it on the touchscreen. 
For details, refer to descriptions in "Parameter Setting". 
Table 2-28 Normalization Reference Level 
Parameter 
Remarks 
Default 
10 dB 
Range 
-200 dB to 200 dB 
Unit 
dB 
Knob Step 
1 dB 
Left/Right Arrow Key Step 
Up/Down Arrow Key Step 
10 dB 