Instructions
RIGOL         Chapter 5 Functions of the Front Panel of VSA 
5-8            RSA3000E User Guide     
AMPT 
Sets the amplitude parameters of the analyzer. You can modify these parameters to 
make the signals under test be displayed with minimal errors in the current window, 
easy for you to observe. 
Auto Scale 
Sets the Y-axis reference level and the Y-axis scale value automatically to ensure that 
the signal can be fully displayed for better observation of the trace. 
Ref Value 
Sets the Y-axis reference of the selected trace. 
Remarks: 
  When "Log Mag" or "Linear Mag" is selected for the trace format, the reference 
level is located at the top of the Y axis; for other conditions, the reference level 
is located in the middle of the Y axis. 
  The reference level is displayed at the top line of the graticule. 
Table 5-9 Reference Level 
Parameter 
Remarks 
Default 
0 
Range 
(-1e+12) to (1e+12) 
Unit 
refer to "Y Axis Unit" 
Knob Step 
1 
Left/Right Arrow Key Step 
Up/Down Arrow Key Step 
10 
Scale/Div 
Sets the Y scale per division of the selected trace. The scale value is displayed at the 
top line of the graticule. 
Note: When you select a different trace format, the default value and unit of the 
Y-axis scale are different. 
Range 
Sets the amplitude of the largest sinusoidal input signal without being clipped by the 
IF ADC. 