Instructions
RIGOL         Chapter 4 Functions of the Front Panel of EMI 
4-20              RSA3000E User Guide  
)
k
10101)(k
10log
(
NewAvg
10
Newdata
10
OldAvg
+×−
=
   (4-9) 
In the above formula, the parameter unit is dB. 
  Scalar: In this mode, all filtering and averaging processes work on the 
voltage envelope of the signal. This average type is the most appropriate 
one for observing the great envelope fluctuations of AM or pulse-modulated 
signals such as radar and TDMA transmitters. The formula is shown as 
follows: 
)(
k
10101)(k
20logNewAvg
20
Newdata
20
OldAvg
+×−
=
   (4-10) 
In the above formula, the parameter unit is dB. 
3.  Avg Auto 
Sets the status of auto average. When the auto average function is enabled, the 
instrument will select the best average type based on the current settings. When 
you select one of the average types manually, the instrument will apply the 
selected type, and the auto average status is set to Manual. 
Limit 
Sets the parameters of limit lines. After you press Preset, the limit line 
measurement function is disabled, but the data of the limit lines will be reserved. The 
limit line data will only be deleted when the loading mode is 
Default. When you exit 
the measurement mode, the limit line data will not be deleted. 
1.  Test Limits 
Selects whether the displayed traces are tested against the displayed limit lines. 
For each displayed trace, the corresponding limit line is turned on, and a 
message will be displayed at the upper-left corner of the trace to indicate 
whether the test passes or fails. 
2.  Select Limit 
Selects the current limit line. By default, it is Limit1. 
3.  Limit State 
Enables or disables the display of the limit line. When the limit line is on, the 
measurement interface displays the limit line, and the corresponding traces are 
tested against the current limit lines. Each limit line is displayed in a different 
color. 
4.  Edit Limit 
When "Display State" of the limit line is set to "On", this menu is valid. When you 