Instructions
Chapter 4 Functions of the Front Panel of EMI  RIGOL 
RSA3000E User Guide  4-3 
Midspan Freq 
Sets the center frequency of the current channel. 
Remarks: 
  When you modify the midspan frequency, the start and stop frequency will be 
modified automatically if the span remains to be unchanged. 
  Modifying the midspan frequency indicates that the frequency is changed along 
the current channel horizontally, and the adjustable range should be within the 
frequency range specified in the technical specifications of the analyzer. 
  You can use the numeric keys, the knob, or arrow keys to modify this parameter; 
you can also use the touch screen to modify the parameter. 
Table 4-2 Midspan Freq 
Parameter 
Remarks 
Default 
515 MHz 
Range 
5 Hz to (Fmax - 5 Hz) 
Unit 
GHz, MHz, kHz, Hz 
Knob Step 
step = RBW (meter)/2 
Left/Right Arrow Key Step 
Up/Down Arrow Key Step 
step = RBW (meter)*10 
Start Freq 
Sets the start frequency of the current frequency channel.  
Remarks: 
  When you modify the start frequency, the span and midspan frequency will be 
changed. 
  You can use the numeric keys, the knob, or arrow keys to modify this parameter; 
you can also use the touch screen to modify the parameter. 
Table 4-3 Start Frequency 
Parameter 
Remarks 
Default 
midspan frequency - span/2 
Range 
0 Hz to (Fmax – 10 Hz) 
Unit 
GHz, MHz, kHz, Hz 
Knob Step 
step = RBW (meter)/2 
Left/Right Arrow Key Step 
Up/Down Arrow Key Step 
step = RBW (meter)*10 