Instructions
Chapter 3 Functions of the Front Panel of RTSA  RIGOL 
RSA3000E User Guide  3-5 
AMPT 
Press AMPT on the front panel to enter the amplitude setting menu. In this menu, 
Reference Level is, by default, selected. 
In non-PvT measurement mode and the PvT measurement mode (except the Ref 
Level and Scale/Div menus), for details about each item under this menu, refer to 
"AMPT" in Chapter 2. 
Ref Level (PvT) 
In the PvT measurement view, adjusting the reference level only changes the 
position of the trace displayed on the screen, but will not cause you to restart 
measurement. Note that this menu is only valid in PvT measurement mode. 
Table 3-4 Ref Level (PvT) 
Parameter 
Remarks 
Default 
0 dBm 
Range 
-250 dBm to 250 dBm 
Unit 
dBm, -dBm, V, mV, μV 
Knob Step 
For "Log" scale type, step = reference level/10 
For "Lin" scale type, step = 0.1 dBm 
Left/Right Arrow Key Step 
Up/Down Arrow Key Step 
For "Log" scale type, step = reference level  
For "Lin" scale type, step = 1 dBm 
Scale/Div (PvT) 
Sets the unit per division in the vertical axis of the PvT view. Note that this menu is 
only valid in PvT measurement mode. 
Table 3-5 Scale/Div (PvT) 
Parameter 
Remarks 
Default 
10 dB 
Range 
0.1 dB to 20 dB 
Unit 
dB 
Knob Step 
scale ≥0.1, step = 1 dB 
scale < 0.1, step = 0.1 dB 
Left/Right Arrow Key Step 
Up/Down Arrow Key Step 
at 1-2-5 step 