Instructions
Chapter 2 Functions of the Front Panel of GPSA  RIGOL 
RSA3000E User Guide  2-81 
Pk-Pk Search 
Executes Peak Search and Minimum Peak functions at the same time, and marks the 
results with the Delta marker. Wherein, the result of Peak Search is marked with the 
reference marker and the result of Minimum Peak is marked with the Delta marker. 
Cont Peak 
Enables or disables continuous peak search. By default, it is Off. When it is enabled, 
after finishing each sweep, the analyzer will automatically execute one peak search 
operation to track the measurement signal. 
When the marker is a fixed type, the Cont Peak menu is grayed out and disabled. The 
signal track and continuous peak functions are mutually exclusive. Once either of 
them is enabled, the menu of the other one will be grayed out. When the Cont Peak 
function is on, if the current marker is off, set the marker to Position mode, and then 
execute the peak search operation. 
Difference between Cont Peak and Signal Track 
In Cont Peak, the system always searches for the maximum in the current 
frequency channel; while in Signal Track, the system will search for and mark the 
point (with no more than 3 dB variation in amplitude) near the marker before 
Signal Track is enabled, and then set the frequency of this point to be the center 
frequency. 
Peak Config 
1.  Peak Threshold 
Specifies the minimum value of the peak amplitude. Only when the peak is 
greater than the peak threshold, can it be judged as a peak. 
Table 2-59 Peak Threshold 
Parameter 
Remarks 
Default 
-90 dBm 
Range 
-200 dBm to 0 dBm 
Unit 
dBm, -dBm, V, mV, uV 
Knob Step 
1 dBm 
Left/Right Arrow Key Step 
Up/Down Arrow Key Step 
5 dBm 
2.  Threshold State 
Enables or disables the peak threshold function. 