Datasheet
22
RIGOL
Humidity Range
Operating
below +30℃: ≤95% RH (without condensation)
+30℃ to +40℃, ≤75% RH (without condensation)
+40℃ to +50℃, ≤45% RH (without condensation)
Non-operating below 65℃: ≤95% RH (without condensation)
Altitude
Operating below 3,000
Non-operating below 15,000
Warranty and Calibration Interval
Warranty and Calibration Interval
Warranty Warranty
Recommended Calibration Interval Recommended Calibration Interval
Regulations
Regulations
Electromagnetic
Compatibility
Compliant with EMC DIRECTIVE 2014/30/EU, compliant with or higher than the standards specified in IEC 61326-
1:2013/EN 61326-1:2013 Group 1 Class A
CISPR 11/EN 55011
IEC 61000-4-2:2008/EN 61000-4-2 ±4.0 kV (contact discharge), ±8.0 kV (air discharge)
IEC 61000-4-3:2002/EN 61000-4-3
3 V/m (80 MHz to 1 GHz); 3 V/m (1.4 GHz to 2 GHz); 1 V/m (2.0 GHz to 2.7
GHz)
IEC 61000-4-4:2004/EN 61000-4-4 1 kV power line
IEC 61000-4-5:2001/EN 61000-4-5
0.5 kV (phase-to-neutral voltage); 1 kV (phase-to-earth voltage); 1 kV
(neutral-to-earth voltage)
IEC 61000-4-6:2003/EN 61000-4-6 3 V, 0.15-80 MHz
IEC 61000-4-11:2004/EN 61000-4-11
voltage dip: 0% UT during half cycle; 0% UT during 1 cycle; 70% UT during
25 cycles
short interruption: 0% UT during 250 cycles
Safety
IEC 61010-1:2010 (Third Edition)/EN 61010-1:2010,
UL 61010-1:2012 R4.16 and CAN/CSA-C22.2 NO. 61010-1-12+ GI1+ GI2
Vibration
Meets GB/T 6587; class 2 random
Meets MIL-PRF-28800F and IEC60068-2-6; class 3 random
Shock
Meets GB/T 6587-2012; class 2 random
Meets MIL-PRF-28800F and IEC60068-2-27; class 3 random
(in non-operating conditions: 30 g, half sine, 11 ms duration, 3 vibrations along the main axis, a total of 18 vibrations)
Mechanical Characteristics
Mechanical Characteristics
Dimensions
[3]
410 mm (W)×224 mm (H)×135 mm (D)
Weight
[4]
Package Excluded <3.9 kg
Package Included <7.1 kg
Rack Mount Kit 6U
Non-volatile Memory
Non-volatile Memory
Data/File Storage
Setup/Image setup (*.stp), image (*.png, *.bmp, *.tif, *.jpg)
Waveform
Data
CSV waveform data (*.csv), binary waveform data (*.bin, *.wfm), list data (*.csv), reference
waveform data (*.ref, *.csv, *.bin), arbitrary waveform data (*.arb)
Reference Waveform Display 10 internal waveforms, and its storage is limited by the capacity
Setting Storage is limited by the capacity
USB Capacity Support the USB storage device that conforms to the industry standard
Note[1]: Maximum value. single-channel, 10 ns horizontal time base, input amplitude 4 div, sine wave signal with 10 MHz frequency. Others are
default settings
Note[2]: 1 mV/div and 2 mV/div are a magnification of 4 mV/div setting. For vertical accuracy calculations, use full scale of 32 mV for 1 mV/div and 2
mV/div sensitivity setting
Note[3]: Supporting legs and handle folded, knob height included, front protective cover excluded
Note[4]: MSO7000 model, standard configuration