Data Sheet

26
RIGOL
Note[1]: Half-channel mode: CH1 and CH2 are considered as a group; CH3 and CH4 are considered as another group. Each group share the same ADC sample,
and either one of the channels in each group is enabled.
Note[2]: Maximum value. single-channel, 10 ns horizontal time base, input amplitude 4 div, sine wave signal with 10 MHz frequency. Others are default settings.
Note[3]: 1 mV/div and 2 mV/div are a magnification of 4 mV/div setting. For vertical accuracy calculations, use full scale of 32 mV for 1 mV/div and 2 mV/div sensitivity setting.
Note[4]: Supporting legs and handle folded, knob height included, front protective cover excluded.
Note[5]: Standard configuration.
Warranty and Calibration Interval
Warranty and Calibration Interval
Warranty 3 years
Recommended Calibration Interval 18 months
Regulations
Regulations
Electromagnetic
Compatibility
Compliant with EMC DIRECTIVE 2014/30/EU, compliant with or higher than the standards specified in IEC 61326-
1:2013/EN 61326-1:2013 Group 1 Class A
CISPR 11/EN 55011
IEC 61000-4-2:2008/EN 61000-4-2 ±4.0 kV (contact discharge), ±8.0 kV (air discharge)
IEC 61000-4-3:2002/EN 61000-4-3
3 V/m (80 MHz to 1 GHz); 3 V/m (1.4 GHz to 2 GHz); 1 V/m (2.0 GHz to 2.7
GHz)
IEC 61000-4-4:2004/EN 61000-4-4 1 kV power line
IEC 61000-4-5:2001/EN 61000-4-5
0.5 kV (phase-to-neutral voltage); 1 kV (phase-to-earth voltage); 1 kV
(neutral-to-earth voltage)
IEC 61000-4-6:2003/EN 61000-4-6 3 V, 0.15-80 MHz
IEC 61000-4-11:2004/EN 61000-4-
11
voltage dip: 0% UT during half cycle; 0% UT during 1 cycle; 70% UT during
25 cycles
short interruption: 0% UT during 250 cycles
Safety
IEC 61010-1:2010 (Third Edition)/EN 61010-1:2010,
UL 61010-1:2012 R4.16 and CAN/CSA-C22.2 NO. 61010-1-12+ GI1+ GI2
Vibration
Meet GB/T 6587; class 2 random
Meets MIL-PRF-28800F and IEC60068-2-6; class 3 random
Shock
Meet GB/T 6587-2012; class 2 random
Meet MIL-PRF-28800F and IEC60068-2-27; class 3 random
(in non-operating conditions: 30 g, half sine, 11 ms duration, 3 vibrations along the main axis, a total of 18 vibrations)
Mechanical Characteristics
Mechanical Characteristics
Dimensions
[4]
367 mm (W)×200 mm (H)×130 mm (D)
Weight
[5]
Package
Excluded
<3.5 kg
Package
Included
<5.8 kg
Rack Mount Kit 5U
Non-volatile Memory
Non-volatile Memory
Data/File Storage
Setup/Image setup (*.stp), image (*.png, *.bmp, *.tif, *.jpg)
Waveform
Data
CSV waveform data (*.csv), binary waveform data (*.bin, *.wfm), list data (*.csv), reference
waveform data (*.ref, *.csv, *.bin), and arbitrary waveform data (*.arb)
Reference Waveform Display 10 internal waveforms, and its storage is limited by the capacity
Setting Storage is limited by the capacity
USB Capacity Support the USB storage device that conforms to the industry standard