RIGOL Chapter 2 Command System
2-228 MSO5000 Programming Guide
:WAVeform:YINCrement?
Syntax
:WAVeform:YINCrement?
Description
Queries the unit voltage of the current channel in the Y axis. Its unit is the same as that of
the current amplitude.
Remarks
The returned value is related to the current data reading mode:
In NORMal mode, YINCrement = VerticalScale/25.
In RAW mode, YINCrement and VerticalScale of the memory waveforms are related to the
currently selected VerticalScale.
In Max mode, YINCrement = VerticalScale/25 when the instrument is in the RUN state;
YINCrement is related to the VerticalScale of the internal waveform and the currently
selected VerticalScale when the instrument is in the Stop state.
Return
Format
The query returns the unit voltage in scientific notation.
:WAVeform:YORigin?
Syntax
:WAVeform:YORigin?
Description
Queries the vertical offset relative to the vertical reference position of the currently
selected channel source in the Y direction.
Remarks
The returned value is related to the current data reading mode:
In NORMal mode, YORigin = VerticalOffset/YINCrement.
In RAW mode, YORigin and VerticalScale of the memory waveforms are related to the
currently selected VerticalScale.
In Max mode, YORigin = VerticalOffset/YINCrement when the instrument is in the RUN
state; YORigin is related to the VerticalScale of the internal waveform and the currently
selected VerticalScale when the instrument is in the Stop state.
Return
Format
The query returns an integer.
:WAVeform:YREFerence?
Syntax
:WAVeform:YREFerence?
Description
Queries the vertical reference position of the currently selected channel source in the Y
direction.
Remarks
The returned value is related to the current data reading mode.
In NORMal mode, YREFerence is a fixed value 128 (screen bottom is 0 and screen top is
255).
In RAW mode, YREFerence is related to the VerticalScale of the memory waveforms and
the currently selected VerticalScale.
In MAX mode, when the instrument is in the RUN state, YREFerence is a fixed value 128
(screen bottom is 0 and screen top is 255); when the instrument is in the STOP state,
YREFerence is related to the VerticalScale of the memory waveforms and the currently
selected VerticalScale.
Return
Format
The query returns an integer.