Instructions

Chapter 19 System Utility Function Setting RIGOL
MSO5000 User's Guide 19-9
Self-calibration
The self-calibration program can quickly make the oscilloscope to work in an optimal
state to get the precise measurement results. You can perform self-calibration at any
time, especially when the changes of the ambient temperature reach or above 5.
Make sure that the oscilloscope has been warmed up or operating for more than 30
minutes before the self-calibration.
Disconnect all the input channels, then press Utility System SelfCal, and the
self-calibration interface is displayed, as shown in the following figure.
Figure 19-2 Self-calibration Interface
Press Start, and then the oscilloscope will start to execute the self-calibration
program.
After the self-calibration program is launched, the Exit menu is no longer grayed
out and is enabled. Press Exit to give up the self-calibration operation at any
time.
Press Window continuously to open or close the self-calibration information
window. You can also enable the touch screen to tap the icon at the
upper-right corner of the self-calibration information window to close the
window.
Note: Most of the keys are disabled during the self-calibration.
Auto Config
You can set the parameters for the AUTO menu.
Press Utility More Auto Config, and set the parameters for the AUTO key.