Data Sheet

24
RIGOL
DC Offset
Range ±2.5 V (HighZ), ±1.25 V (50 Ω)
Resolution 100 uV or 3 bits (whichever is greater)
Accuracy ±(2% of offset setting+5 mV+0.5% of amplitude)
Modulation
AM, FM, FSK
AM
Modulating Waveforms: Sine, Square, Triangle, and Noise.
Modulation Frequency: 1 Hz to 50 kHz
Modulation Depth: 0% to 120%
FM
Modulating Waveforms: Sine, Square, Triangle, and Noise.
Modulation Frequency: 1 Hz to 50 kHz
Modulation Offset: 1 Hz to carrier frequency
FSK
Modulating Waveforms: 50% duty cycle square
Modulation Frequency: 1 Hz to 50 kHz
Hopping Frequency: 100 mHz to max. carrier frequency
Sweep
Linear, Log, and Step
Sweep Time 1 ms to 500 s
Start Frequency and End
Frequency
Any frequencies within the waveform range
Burst
N Cycle, Infinite
Cycle Count 1 to 1000000
Burst Period 1 μs to 500 s
Burst Delay 0 s to 100 s
Trigger Source Internal, Manual
Digital Voltmeter
Digital Voltmeter (technical specifications are typical values)
Source Any analog channel
Function DC, AC+DC RMS, and AC RMS
Resolution ACV/DCV: 3 bits
Limits Beeper Sound an alarm when the voltage value is within or outside of the limit range.
Range Measurement
Display the latest measurement results in the form of a diagram, and display the extrema over the
last 3 seconds
High-precision Frequency Counter
High-precision Frequency Counter
Source Any analog channel and digital channel
Measure Frequency, period, totalizer
Counter
Resolution Max. 6 bits, user-defined
Max. Frequency Max. bandwidth of the analog channel
Totalizer
48-bit totalizer
Edge Count the number of the rising edges
Time Reference Internal Reference
Customization for Quick Key
Customization for Quick Key
Quick Screenshot
Quickly save the screen image to the specified path based on the current image storage menu
settings.
Quick Waveform Save
Quickly save the screen or memory waveforms to the specified path based on the current
waveform storage menu settings.
Quick Save Settings Quickly save the setup file to the specified path based on the current setup storage menu settings.
Quick All Measurement Display all the prompt message windows for all the measurement of the waveforms.
Quick Reset of Statistics
Quickly reset all the measurement statistics data and measurement counts.
Quickly reset all the statistics information in PassFail function.
Quick Waveform Recording Quickly start or stop the waveform recording.
Quick Email Sending Quickly send the Email based on the set email address.