Instructions
Table Of Contents
- Guaranty and Declaration
- Safety Requirement
- MSO5000-E Series Overview
- Document Overview
- Quick Start
- General Inspection
- Appearance and Dimensions
- To Prepare for Use
- Front Panel Overview
- Rear Panel Overview
- Front Panel Function Overview
- User Interface
- Touch Screen Controls
- Parameter Setting Method
- To Use the Kensington Security Lock
- To Use the Built-in Help System
- To View the Option Information and the Option Installation
- To Set the Vertical System
- To Set the Horizontal System
- To Set the Sample System
- To Trigger the Oscilloscope
- Trigger Source
- Trigger LEVEL/Threshold Level
- Trigger Mode
- Trigger Coupling
- Trigger Holdoff
- Noise Rejection
- Trigger Type
- Edge Trigger
- Pulse Trigger
- Slope Trigger
- Video Trigger
- Pattern Trigger
- Duration Trigger
- Timeout Trigger
- Runt Trigger
- Window Trigger
- Delay Trigger
- Setup/Hold Trigger
- Nth Edge Trigger
- RS232 Trigger (Option)
- I2C Trigger (Option)
- SPI Trigger (Option)
- CAN Trigger (Option)
- FlexRay Trigger (Option)
- LIN Trigger (Option)
- I2S Trigger (Option)
- MIL-STD-1553 Trigger (Option)
- Zone Trigger
- Trigger Output Connector
- Operations and Measurements
- Digital Voltmeter (DVM) and Frequency Counter
- Power Analysis (Option)
- Histogram Analysis
- Digital Channel
- Protocol Decoding
- Reference Waveform
- To Enable Ref Function
- To Select the Reference Channel
- To Select the Ref Source
- To Adjust the Ref Waveform Display
- To Save to Internal Memory
- To Clear the Display of the Reference Waveform
- To View Details of the Reference Waveform
- To Reset the Reference Waveform
- Color Setting
- Label Setting
- To Export to Internal or External Memory
- To Import from Internal or External Memory
- Pass/Fail Test
- Waveform Recording & Playing
- Search and Navigation Function
- Display Control
- Function/Arbitrary Waveform Generator (Option)
- Store and Load
- System Utility Function Setting
- Remote Control
- Troubleshooting
- Appendix
- Index

Chapter 5 To Trigger the Oscilloscope RIGOL
MSO5000-E User Guide 5-11
Polarity:
Press Polarity continuously to select the desired polarity. The polarities available are
positive polarity (
) and negative polarity ( ).
Trigger Condition:
Press When, then rotate the multifunction knob to select the trigger condition,
and then press down the knob to select the trigger condition. You can also press
When continuously to select the trigger condition or enable the touch screen to tap
the desired trigger condition and select it.
When you select "Positive" for polarity, ">" for trigger condition, the oscilloscope
triggers when the positive pulse width of the input signal is greater than the
specified pulse width.
When you select "Positive" for polarity, "<" for trigger condition, the oscilloscope
triggers when the positive pulse width of the input signal is smaller than the
specified pulse width.
When you select "Positive" for polarity, "<>" for trigger condition, the
oscilloscope triggers when the positive pulse width of the input signal is greater
than the specified lower limit of pulse width and smaller than the specified
upper limit of pulse width.
When you select "Negative" for polarity, ">" for trigger condition, the
oscilloscope triggers when the negative pulse width of the input signal is greater
than the specified pulse width.
When you select "Negative" for polarity, "<" for trigger condition, the
oscilloscope triggers when the negative pulse width of the input signal is smaller
than the specified pulse width.
When you select "Negative" for polarity, "<>" for trigger condition, the
oscilloscope triggers when the negative pulse width of the input signal is greater
than the specified lower limit of pulse width and smaller than the specified
upper limit of pulse width.
Pulse Width Setting:
When ">" or "<" is set to the trigger condition, press Lower or Upper, and
then rotate the multifunction knob to set the desired value; or press down the
multifunction knob
to set the lower limit value or the upper limit value. You
can also use the pop-up numeric keypad to set it. The pulse range available is
from 800 ps to 10 s.
When "<>" is set to the trigger condition, press Upper and Lower, and then
rotate the multifunction knob
to set the lower limit value and the upper limit
value. You can also use the pop-up numeric keypad to set them.
Note: The lower limit of the pulse width must be smaller than the upper limit.
Trigger Mode:
In the trigger control area (Trigger) on the front panel, press Mode to quickly switch