Instructions
Table Of Contents
- Guaranty and Declaration
- Safety Requirement
- MSO5000-E Series Overview
- Document Overview
- Quick Start
- General Inspection
- Appearance and Dimensions
- To Prepare for Use
- Front Panel Overview
- Rear Panel Overview
- Front Panel Function Overview
- User Interface
- Touch Screen Controls
- Parameter Setting Method
- To Use the Kensington Security Lock
- To Use the Built-in Help System
- To View the Option Information and the Option Installation
- To Set the Vertical System
- To Set the Horizontal System
- To Set the Sample System
- To Trigger the Oscilloscope
- Trigger Source
- Trigger LEVEL/Threshold Level
- Trigger Mode
- Trigger Coupling
- Trigger Holdoff
- Noise Rejection
- Trigger Type
- Edge Trigger
- Pulse Trigger
- Slope Trigger
- Video Trigger
- Pattern Trigger
- Duration Trigger
- Timeout Trigger
- Runt Trigger
- Window Trigger
- Delay Trigger
- Setup/Hold Trigger
- Nth Edge Trigger
- RS232 Trigger (Option)
- I2C Trigger (Option)
- SPI Trigger (Option)
- CAN Trigger (Option)
- FlexRay Trigger (Option)
- LIN Trigger (Option)
- I2S Trigger (Option)
- MIL-STD-1553 Trigger (Option)
- Zone Trigger
- Trigger Output Connector
- Operations and Measurements
- Digital Voltmeter (DVM) and Frequency Counter
- Power Analysis (Option)
- Histogram Analysis
- Digital Channel
- Protocol Decoding
- Reference Waveform
- To Enable Ref Function
- To Select the Reference Channel
- To Select the Ref Source
- To Adjust the Ref Waveform Display
- To Save to Internal Memory
- To Clear the Display of the Reference Waveform
- To View Details of the Reference Waveform
- To Reset the Reference Waveform
- Color Setting
- Label Setting
- To Export to Internal or External Memory
- To Import from Internal or External Memory
- Pass/Fail Test
- Waveform Recording & Playing
- Search and Navigation Function
- Display Control
- Function/Arbitrary Waveform Generator (Option)
- Store and Load
- System Utility Function Setting
- Remote Control
- Troubleshooting
- Appendix
- Index

RIGOL Chapter 12 Reference Waveform
12-2 MSO5000-E User Guide
To Enable Ref Function
Press Ref on the front panel to enable the reference waveform function, or enable
the touch screen and then tap the function navigation icon at the lower-left
corner of the screen to open the function navigation. Then, tap the "Ref" icon to
enable the reference waveform function.
When the Ref function is enabled, you can select a different color for each reference
waveform, set the source of each reference channel, adjust the vertical scale and
offset of the reference waveform, save the reference waveform to the internal or
external memory, and recall it when needed.
Note: When the horizontal timebase is set to XY, the reference waveform function is
disabled.
To Select the Reference Channel
Press Current, rotate the multifunction knob to select the desired reference
waveform channel (Ref1-Ref10). You can also press Current continuously or enable
the touch screen to select the reference waveform channel. By default, Ref1 is
enabled.
To Select the Ref Source
Press Source, rotate the multifunction knob to select the desired reference
waveform source (CH1-CH2, D0-D15, or Math1-Math4). You can also press Source
continuously or enable the touch screen to select the reference waveform source.
Note: Only the currently enabled channel can be selected as the source channel of
the reference waveform.
To Adjust the Ref Waveform Display
After pressing SaveToRef, you can adjust the vertical scale and offset of the
reference waveform specified under Current.
Press VScale, then rotate the multifunction knob or use the pop-up
numeric keypad to set the vertical scale of the reference waveform. You can also
enable the touch screen and use the "Pinch & Stretch" gesture to adjust the
vertical scale.