Instructions
Table Of Contents
- Guaranty and Declaration
- Safety Requirement
- MSO5000-E Series Overview
- Document Overview
- Quick Start
- General Inspection
- Appearance and Dimensions
- To Prepare for Use
- Front Panel Overview
- Rear Panel Overview
- Front Panel Function Overview
- User Interface
- Touch Screen Controls
- Parameter Setting Method
- To Use the Kensington Security Lock
- To Use the Built-in Help System
- To View the Option Information and the Option Installation
- To Set the Vertical System
- To Set the Horizontal System
- To Set the Sample System
- To Trigger the Oscilloscope
- Trigger Source
- Trigger LEVEL/Threshold Level
- Trigger Mode
- Trigger Coupling
- Trigger Holdoff
- Noise Rejection
- Trigger Type
- Edge Trigger
- Pulse Trigger
- Slope Trigger
- Video Trigger
- Pattern Trigger
- Duration Trigger
- Timeout Trigger
- Runt Trigger
- Window Trigger
- Delay Trigger
- Setup/Hold Trigger
- Nth Edge Trigger
- RS232 Trigger (Option)
- I2C Trigger (Option)
- SPI Trigger (Option)
- CAN Trigger (Option)
- FlexRay Trigger (Option)
- LIN Trigger (Option)
- I2S Trigger (Option)
- MIL-STD-1553 Trigger (Option)
- Zone Trigger
- Trigger Output Connector
- Operations and Measurements
- Digital Voltmeter (DVM) and Frequency Counter
- Power Analysis (Option)
- Histogram Analysis
- Digital Channel
- Protocol Decoding
- Reference Waveform
- To Enable Ref Function
- To Select the Reference Channel
- To Select the Ref Source
- To Adjust the Ref Waveform Display
- To Save to Internal Memory
- To Clear the Display of the Reference Waveform
- To View Details of the Reference Waveform
- To Reset the Reference Waveform
- Color Setting
- Label Setting
- To Export to Internal or External Memory
- To Import from Internal or External Memory
- Pass/Fail Test
- Waveform Recording & Playing
- Search and Navigation Function
- Display Control
- Function/Arbitrary Waveform Generator (Option)
- Store and Load
- System Utility Function Setting
- Remote Control
- Troubleshooting
- Appendix
- Index

Chapter 11 Protocol Decoding RIGOL
MSO5000-E User Guide 11-19
2. Quickly apply SPI trigger settings to SPI decoding
Press Copy Trig to copy the SPI trigger settings and apply them to SPI
decoding function (set the corresponding SPI decoding parameters
automatically). For analog channels, the threshold level settings will also be
copied.
3. Mode
Press Mode to enter the mode setting menu. Press Mode continuously to select
"Timeout" or "CS".
Timeout: you can perform frame synchronization according to the timeout,
and the timeout value must be greater than half of the clock cycle. Press
Timeout, and then rotate the multifunction knob or use the pop-up
numeric keypad to set the timeout value. The available range of the timeout
value is from 8 ns to 10 s. By default, it is 1 μs.
CS: contains a chip select line (CS). You can perform frame synchronization
according to CS. When "CS" is selected,
Press CS and rotate the multifunction knob to select the desired
channel, and then press down the knob to select it. You can also press
CS continuously or enable the touch screen to select it. The analog
channel (CH1-CH2) and digital channel (D0-D15) can all be selected as
the CS channel.
Press CS Polarity continuously to select (Positive) or
(Negative) as the polarity for CS channel.
When you select the analog channel (CH1-CH2) as the CS channel,
press Threshold, then rotate the multifunction knob or use the
pop-up numeric keypad to set the threshold.
4. Source setting
Press Sources to enter the source setting menu.
CLK setting
Press CLK and rotate the multifunction knob to select the desired
source of the clock channel, and then press down the knob to select it.
You can also press CLK continuously or enable the touch screen to
select it. The analog channel (CH1-CH2) and digital channel (D0-D15)
can all be selected.
When you select the analog channel (CH1-CH2), press Threshold
below the CLK menu, then rotate the multifunction knob or use
the pop-up numeric keypad to set the threshold of the clock channel.
Press Slope continuously to set the instrument to sample MISO and
MOSI on the rising edge (
) or falling edge ( ) of the clock signal.
MISO and MOSI setting
Press MISO and rotate the multifunction knob to select the
desired MISO data channel, and then press down the knob to select it.