Instructions
Table Of Contents
- Guaranty and Declaration
- Safety Requirement
- MSO5000-E Series Overview
- Document Overview
- Quick Start
- General Inspection
- Appearance and Dimensions
- To Prepare for Use
- Front Panel Overview
- Rear Panel Overview
- Front Panel Function Overview
- User Interface
- Touch Screen Controls
- Parameter Setting Method
- To Use the Kensington Security Lock
- To Use the Built-in Help System
- To View the Option Information and the Option Installation
- To Set the Vertical System
- To Set the Horizontal System
- To Set the Sample System
- To Trigger the Oscilloscope
- Trigger Source
- Trigger LEVEL/Threshold Level
- Trigger Mode
- Trigger Coupling
- Trigger Holdoff
- Noise Rejection
- Trigger Type
- Edge Trigger
- Pulse Trigger
- Slope Trigger
- Video Trigger
- Pattern Trigger
- Duration Trigger
- Timeout Trigger
- Runt Trigger
- Window Trigger
- Delay Trigger
- Setup/Hold Trigger
- Nth Edge Trigger
- RS232 Trigger (Option)
- I2C Trigger (Option)
- SPI Trigger (Option)
- CAN Trigger (Option)
- FlexRay Trigger (Option)
- LIN Trigger (Option)
- I2S Trigger (Option)
- MIL-STD-1553 Trigger (Option)
- Zone Trigger
- Trigger Output Connector
- Operations and Measurements
- Digital Voltmeter (DVM) and Frequency Counter
- Power Analysis (Option)
- Histogram Analysis
- Digital Channel
- Protocol Decoding
- Reference Waveform
- To Enable Ref Function
- To Select the Reference Channel
- To Select the Ref Source
- To Adjust the Ref Waveform Display
- To Save to Internal Memory
- To Clear the Display of the Reference Waveform
- To View Details of the Reference Waveform
- To Reset the Reference Waveform
- Color Setting
- Label Setting
- To Export to Internal or External Memory
- To Import from Internal or External Memory
- Pass/Fail Test
- Waveform Recording & Playing
- Search and Navigation Function
- Display Control
- Function/Arbitrary Waveform Generator (Option)
- Store and Load
- System Utility Function Setting
- Remote Control
- Troubleshooting
- Appendix
- Index

RIGOL Chapter 11 Protocol Decoding
11-12 MSO5000-E User Guide
the top of the event table to select it.
If you select "Packets", the time and decoded data are displayed in the
event table; if you select "Details", the detailed data in the specified
row will be displayed in the event table; if you select "Payload", all data
in the specified column will be displayed in the event table.
Export: If you select "Packets", you can export time, the corresponding
decoded data, and etc. Press Export, the save setting menu is
displayed. You can export the data table of the packets to the internal
or external USB storage device (when detected) in CSV format. For
detailed saving operation, refer to descriptions in "
Store and Load".
Jump to: Set the operating status of the oscilloscope to "STOP". Then,
select the data of the specified time from the event table. Press Jump
to and the specified data waveforms are displayed in the center of the
screen.
Decode: The setting of this menu enables you to quickly view the
corresponding event tables of four decode modules (Decode1,
Decode2, Decode3, and Decode4). Press Decode and rotate the
multifunction knob
to select a decode module, then press down
the knob to select it. You can also press Decode continuously or
enable the touch screen to select it.
7. Error expressions in decoding
In RS232 decoding, parity error and/or EOF (end of frame) error may occur.
Parity Error
When a parity error is detected during the decoding, red error report
information will be displayed. For example, when "none" is chosen for the
parity on the transmitter and "odd" is chosen for the parity on the decoder,
the following parity error information
(its display form is related to the
value of the horizontal timebase) occurs. Wherein, the number of 1s in the
8-digit figure (00000100) is an odd number, and the parity bit should be 0.
However, the parity bit detected on the Tx is 1; thus a parity error occurs.