Instructions
Table Of Contents
- Guaranty and Declaration
- Safety Requirement
- MSO5000-E Series Overview
- Document Overview
- Quick Start
- General Inspection
- Appearance and Dimensions
- To Prepare for Use
- Front Panel Overview
- Rear Panel Overview
- Front Panel Function Overview
- User Interface
- Touch Screen Controls
- Parameter Setting Method
- To Use the Kensington Security Lock
- To Use the Built-in Help System
- To View the Option Information and the Option Installation
- To Set the Vertical System
- To Set the Horizontal System
- To Set the Sample System
- To Trigger the Oscilloscope
- Trigger Source
- Trigger LEVEL/Threshold Level
- Trigger Mode
- Trigger Coupling
- Trigger Holdoff
- Noise Rejection
- Trigger Type
- Edge Trigger
- Pulse Trigger
- Slope Trigger
- Video Trigger
- Pattern Trigger
- Duration Trigger
- Timeout Trigger
- Runt Trigger
- Window Trigger
- Delay Trigger
- Setup/Hold Trigger
- Nth Edge Trigger
- RS232 Trigger (Option)
- I2C Trigger (Option)
- SPI Trigger (Option)
- CAN Trigger (Option)
- FlexRay Trigger (Option)
- LIN Trigger (Option)
- I2S Trigger (Option)
- MIL-STD-1553 Trigger (Option)
- Zone Trigger
- Trigger Output Connector
- Operations and Measurements
- Digital Voltmeter (DVM) and Frequency Counter
- Power Analysis (Option)
- Histogram Analysis
- Digital Channel
- Protocol Decoding
- Reference Waveform
- To Enable Ref Function
- To Select the Reference Channel
- To Select the Ref Source
- To Adjust the Ref Waveform Display
- To Save to Internal Memory
- To Clear the Display of the Reference Waveform
- To View Details of the Reference Waveform
- To Reset the Reference Waveform
- Color Setting
- Label Setting
- To Export to Internal or External Memory
- To Import from Internal or External Memory
- Pass/Fail Test
- Waveform Recording & Playing
- Search and Navigation Function
- Display Control
- Function/Arbitrary Waveform Generator (Option)
- Store and Load
- System Utility Function Setting
- Remote Control
- Troubleshooting
- Appendix
- Index

Chapter 10 Digital Channel RIGOL
MSO5000-E User Guide 10-3
Enable/disable a single digital channel
Press Dx On/Off to open the channel selection list. Rotate the multifunction
knob
to select a channel, and then press down the knob to enable or disable
it. You can also press Dx On/Off continuously to switch to select any channel,
and press down the multifunction knob
to enable or disable the channel.
You can also enable the touch screen to enable or disable the selected channel.
Besides, you can also enable the touch screen to tap the digital channel status
area at the bottom of the screen. Then, the LA window is displayed. In the
window, you can select the LA channel and set the threshold. Tap a single
channel to enable or disable the channel.
Note: While pressing D7-D0 to disable all the channels D7-D0, you can still
disable or enable any channel among D7-D0 through the Dx On/Off key. While
pressing D7-D0 to enable D7-D0, the D0-D7 channels under Dx On/Off are
enabled automatically. The same is true for D15-D8 channels.
Enable or disable user-defined channel group
Press Group On/Off, then rotate the multifunction knob to select the
specified channel group. Press down the knob to enable or disable all the
channels in this channel group. You can also enable the touch screen to tap the
specified group and select it. The enabled channel groups are marked with
,
and the disabled channel groups are marked with
.
Note: This function is only valid when the channels have been grouped by users.
For how to carry out group settings, refer to descriptions in "Group Setting".
To Set the Threshold and Calibrate Probe
Threshold Setting
The threshold levels of channels D7-D0 and channels D15-D8 can be set
separately according to your needs. Press Threshold to enter the threshold
setting menu. When the voltage of the input signal is greater than the currently
set threshold, it is treated as logic 1; otherwise, it is treated as logic 0.
Press D7-D0 or D15-D8 respectively to open the threshold selection list and
select the preset threshold. The available preset values include TTL(1.40 V),
CMOS5.0(2.5 V), CMOS3.3(1.65 V), CMOS2.5(1.25 V), CMOS1.8(0.9 V),
ECL(-1.3 V), PECL(3.7 V), LVDS(1.2 V), and 0 V. You can also press D7-D0 or
D15-D8 respectively, then rotate the multifunction knob at a small step
value or use the pop-up numeric keypad to set the desired threshold.
Besides, you can also enable the touch screen to tap the digital channel status
area at the bottom of the screen. Then, the LA window is displayed. In the
window, you can select the LA channel and set the threshold. Tap the
icon