Instructions
Table Of Contents
- Guaranty and Declaration
- Safety Requirement
- MSO5000-E Series Overview
- Document Overview
- Quick Start
- General Inspection
- Appearance and Dimensions
- To Prepare for Use
- Front Panel Overview
- Rear Panel Overview
- Front Panel Function Overview
- User Interface
- Touch Screen Controls
- Parameter Setting Method
- To Use the Kensington Security Lock
- To Use the Built-in Help System
- To View the Option Information and the Option Installation
- To Set the Vertical System
- To Set the Horizontal System
- To Set the Sample System
- To Trigger the Oscilloscope
- Trigger Source
- Trigger LEVEL/Threshold Level
- Trigger Mode
- Trigger Coupling
- Trigger Holdoff
- Noise Rejection
- Trigger Type
- Edge Trigger
- Pulse Trigger
- Slope Trigger
- Video Trigger
- Pattern Trigger
- Duration Trigger
- Timeout Trigger
- Runt Trigger
- Window Trigger
- Delay Trigger
- Setup/Hold Trigger
- Nth Edge Trigger
- RS232 Trigger (Option)
- I2C Trigger (Option)
- SPI Trigger (Option)
- CAN Trigger (Option)
- FlexRay Trigger (Option)
- LIN Trigger (Option)
- I2S Trigger (Option)
- MIL-STD-1553 Trigger (Option)
- Zone Trigger
- Trigger Output Connector
- Operations and Measurements
- Digital Voltmeter (DVM) and Frequency Counter
- Power Analysis (Option)
- Histogram Analysis
- Digital Channel
- Protocol Decoding
- Reference Waveform
- To Enable Ref Function
- To Select the Reference Channel
- To Select the Ref Source
- To Adjust the Ref Waveform Display
- To Save to Internal Memory
- To Clear the Display of the Reference Waveform
- To View Details of the Reference Waveform
- To Reset the Reference Waveform
- Color Setting
- Label Setting
- To Export to Internal or External Memory
- To Import from Internal or External Memory
- Pass/Fail Test
- Waveform Recording & Playing
- Search and Navigation Function
- Display Control
- Function/Arbitrary Waveform Generator (Option)
- Store and Load
- System Utility Function Setting
- Remote Control
- Troubleshooting
- Appendix
- Index

Chapter 9 Histogram Analysis RIGOL
MSO5000-E User Guide 9-3
To Select the Histogram Source
If "Horizontal" or "Vertical" is selected under Type, you need to set the source
(CH1-CH2). Press Source, then rotate the multifunction knob to select the
desired source, and then press down the knob to select the source. You can also
press Source continuously or enable the touch screen to select it.
When you select "Measure" as the histogram type, press Measure on the front
panel to enter the measurement setting menu, and then add one or multiple
measurement items. The added measurement item(s) can be served as the
histogram source(s).
To Set the Histogram Height
The histogram height indicates the number of grids the histogram bar graph should
use on the screen. Press Height, then rotate the multifunction knob or use the
pop-up numeric keypad to set the height. The available range of height is from 1 to 4.
By default, it is 2.
To Set the Histogram Range
When "Horizontal" or "Vertical" is selected under Type, you need to set the window
range. Set "Left Limit", "Right Limit", "Top Limit", and "Bottom Limit" respectively to
adjust the size and position of the histogram window. Press Range to enter the
sub-menu of histogram range. Then press the corresponding softkey and rotate the
multifunction knob
to set their values.
Note: The adjustment for the horizontal timebase and vertical scale will not affect
the timebase of the histogram range, but only shows variation with the scale.
To Enable or Disable the Statistical Function
Press Statistic to enable or disable the statistical function. When the statistical
function is enabled, the statistical results of the histogram data will be displayed, as
shown in the following figure. Note that when you select "Measure" under Type, the
statistical data results also include the parameter XScale.