Instructions
Table Of Contents
- Guaranty and Declaration
- Safety Requirement
- MSO5000-E Series Overview
- Document Overview
- Quick Start
- General Inspection
- Appearance and Dimensions
- To Prepare for Use
- Front Panel Overview
- Rear Panel Overview
- Front Panel Function Overview
- User Interface
- Touch Screen Controls
- Parameter Setting Method
- To Use the Kensington Security Lock
- To Use the Built-in Help System
- To View the Option Information and the Option Installation
- To Set the Vertical System
- To Set the Horizontal System
- To Set the Sample System
- To Trigger the Oscilloscope
- Trigger Source
- Trigger LEVEL/Threshold Level
- Trigger Mode
- Trigger Coupling
- Trigger Holdoff
- Noise Rejection
- Trigger Type
- Edge Trigger
- Pulse Trigger
- Slope Trigger
- Video Trigger
- Pattern Trigger
- Duration Trigger
- Timeout Trigger
- Runt Trigger
- Window Trigger
- Delay Trigger
- Setup/Hold Trigger
- Nth Edge Trigger
- RS232 Trigger (Option)
- I2C Trigger (Option)
- SPI Trigger (Option)
- CAN Trigger (Option)
- FlexRay Trigger (Option)
- LIN Trigger (Option)
- I2S Trigger (Option)
- MIL-STD-1553 Trigger (Option)
- Zone Trigger
- Trigger Output Connector
- Operations and Measurements
- Digital Voltmeter (DVM) and Frequency Counter
- Power Analysis (Option)
- Histogram Analysis
- Digital Channel
- Protocol Decoding
- Reference Waveform
- To Enable Ref Function
- To Select the Reference Channel
- To Select the Ref Source
- To Adjust the Ref Waveform Display
- To Save to Internal Memory
- To Clear the Display of the Reference Waveform
- To View Details of the Reference Waveform
- To Reset the Reference Waveform
- Color Setting
- Label Setting
- To Export to Internal or External Memory
- To Import from Internal or External Memory
- Pass/Fail Test
- Waveform Recording & Playing
- Search and Navigation Function
- Display Control
- Function/Arbitrary Waveform Generator (Option)
- Store and Load
- System Utility Function Setting
- Remote Control
- Troubleshooting
- Appendix
- Index

Chapter 6 Operations and Measurements RIGOL
MSO5000-E User Guide 6-35
reference is negative. The area measured is the algebraic sum of the whole
period area.
Measurement Settings
1. Measurement Category
Press Measure Add Category to select "Horizontal", "Vertical", or
"Other". Meanwhile, the specific measurement parameters are displayed on the
screen, as shown in the figure below. In the measurement category area, you
can also rotate the multifunction knob
to select "Horizontal", "Vertical", or
"Other". Then press down the knob to switch to the desired measurement
category.
"Horizontal" measurement parameters: Period, Freq, RiseTime, FallTime,
+Width, -Width, +Duty, -Duty, +Pulses, -Pulses, +Edges, -Edges, Tvmax,
Tvmin, +SlewRate, and –SlewRate.
"Vertical" measurement parameters: Vmax, Vmin, Vpp, Vtop, Vbase, Vamp,
Vmid, Vlower, Vavg, VRMS, Per.VRMS, Overshot, Preshoot, Area, Per.Area,
and Variance.
"Other" measurement parameters: Delay(A
→B ), Delay(A →B ),
Delay(A
→B ), Delay(A →B ), Phase(A →B ), Phase(A →B ),
Phase(A
→B ), and Phase(A →B ).
In the measurement parameter area, rotate the multifunction knob to select
a measurement item and then press down the knob to select it. Then the
measurement results of the item are displayed at the bottom of the screen
(measurement results of up to 10 items can be displayed at most). When you
add a new measurement item, its measurement result will be displayed as the
first, with the original items moving right in sequence. If the number of
measurement items added exceeds 10, then when you add one item, the
rightmost item at the bottom of the screen will be disabled. You can also enable
the touch screen to select the measurement category and the measurement
Measurement Category Area Measurement Parameter Area