Instructions
Table Of Contents
- Guaranty and Declaration
- Safety Requirement
- MSO5000-E Series Overview
- Document Overview
- Quick Start
- General Inspection
- Appearance and Dimensions
- To Prepare for Use
- Front Panel Overview
- Rear Panel Overview
- Front Panel Function Overview
- User Interface
- Touch Screen Controls
- Parameter Setting Method
- To Use the Kensington Security Lock
- To Use the Built-in Help System
- To View the Option Information and the Option Installation
- To Set the Vertical System
- To Set the Horizontal System
- To Set the Sample System
- To Trigger the Oscilloscope
- Trigger Source
- Trigger LEVEL/Threshold Level
- Trigger Mode
- Trigger Coupling
- Trigger Holdoff
- Noise Rejection
- Trigger Type
- Edge Trigger
- Pulse Trigger
- Slope Trigger
- Video Trigger
- Pattern Trigger
- Duration Trigger
- Timeout Trigger
- Runt Trigger
- Window Trigger
- Delay Trigger
- Setup/Hold Trigger
- Nth Edge Trigger
- RS232 Trigger (Option)
- I2C Trigger (Option)
- SPI Trigger (Option)
- CAN Trigger (Option)
- FlexRay Trigger (Option)
- LIN Trigger (Option)
- I2S Trigger (Option)
- MIL-STD-1553 Trigger (Option)
- Zone Trigger
- Trigger Output Connector
- Operations and Measurements
- Digital Voltmeter (DVM) and Frequency Counter
- Power Analysis (Option)
- Histogram Analysis
- Digital Channel
- Protocol Decoding
- Reference Waveform
- To Enable Ref Function
- To Select the Reference Channel
- To Select the Ref Source
- To Adjust the Ref Waveform Display
- To Save to Internal Memory
- To Clear the Display of the Reference Waveform
- To View Details of the Reference Waveform
- To Reset the Reference Waveform
- Color Setting
- Label Setting
- To Export to Internal or External Memory
- To Import from Internal or External Memory
- Pass/Fail Test
- Waveform Recording & Playing
- Search and Navigation Function
- Display Control
- Function/Arbitrary Waveform Generator (Option)
- Store and Load
- System Utility Function Setting
- Remote Control
- Troubleshooting
- Appendix
- Index

Chapter 6 Operations and Measurements RIGOL
MSO5000-E User Guide 6-33
Note:
Source A and Source B can be any channel among CH1-CH2, Math1-Math4, or
D0-D15. You can press Measure Add to enter the sub-menu to set the
parameters.
The default threshold middle value is 50%. You can press Measure Setting
to enter the sub-menu to set the parameters.
Voltage Parameters
Figure 6-2 Voltage Parameters
1. Vmax: indicates the voltage value from the highest point of the waveform to
the GND.
2. Vmin: indicates the voltage value from the lowest point of the waveform to the
GND.
3. Vpp: the voltage value from the highest point to the lowest point of the
waveform.
4. Vtop: indicates the voltage value from the flat top of the waveform to the GND.
5. Vbase: indicates the voltage value from the flat base of the waveform to the
GND.
6. Vamp: indicates the voltage value from the top of the waveform to the base of
the waveform.
7. Vupper: indicates the actual voltage value that corresponds to the threshold
maximum value.
8. Vmid: indicates the actual voltage value that corresponds to the threshold
middle value.
9. Vlower: indicates the actual voltage value that corresponds to the threshold
minimum value.
10. Vavg: indicates the arithmetic average value on the whole waveform or in the
Preshoot
Overshoot
Vamp
Vbase
Vmax
Vmin
Vpp
Vtop
Vupper
Vmid
Vlower