Instructions
Table Of Contents
- Guaranty and Declaration
- Safety Requirement
- MSO5000-E Series Overview
- Document Overview
- Quick Start
- General Inspection
- Appearance and Dimensions
- To Prepare for Use
- Front Panel Overview
- Rear Panel Overview
- Front Panel Function Overview
- User Interface
- Touch Screen Controls
- Parameter Setting Method
- To Use the Kensington Security Lock
- To Use the Built-in Help System
- To View the Option Information and the Option Installation
- To Set the Vertical System
- To Set the Horizontal System
- To Set the Sample System
- To Trigger the Oscilloscope
- Trigger Source
- Trigger LEVEL/Threshold Level
- Trigger Mode
- Trigger Coupling
- Trigger Holdoff
- Noise Rejection
- Trigger Type
- Edge Trigger
- Pulse Trigger
- Slope Trigger
- Video Trigger
- Pattern Trigger
- Duration Trigger
- Timeout Trigger
- Runt Trigger
- Window Trigger
- Delay Trigger
- Setup/Hold Trigger
- Nth Edge Trigger
- RS232 Trigger (Option)
- I2C Trigger (Option)
- SPI Trigger (Option)
- CAN Trigger (Option)
- FlexRay Trigger (Option)
- LIN Trigger (Option)
- I2S Trigger (Option)
- MIL-STD-1553 Trigger (Option)
- Zone Trigger
- Trigger Output Connector
- Operations and Measurements
- Digital Voltmeter (DVM) and Frequency Counter
- Power Analysis (Option)
- Histogram Analysis
- Digital Channel
- Protocol Decoding
- Reference Waveform
- To Enable Ref Function
- To Select the Reference Channel
- To Select the Ref Source
- To Adjust the Ref Waveform Display
- To Save to Internal Memory
- To Clear the Display of the Reference Waveform
- To View Details of the Reference Waveform
- To Reset the Reference Waveform
- Color Setting
- Label Setting
- To Export to Internal or External Memory
- To Import from Internal or External Memory
- Pass/Fail Test
- Waveform Recording & Playing
- Search and Navigation Function
- Display Control
- Function/Arbitrary Waveform Generator (Option)
- Store and Load
- System Utility Function Setting
- Remote Control
- Troubleshooting
- Appendix
- Index

RIGOL Chapter 6 Operations and Measurements
6-16 MSO5000-E User Guide
Press Offset to set the vertical offset of the operation results. For setting
methods, refer to the descriptions in "
Parameter Setting Method". You can
also enable the touch screen and use the "Drag" gesture to adjust the vertical
offset.
Press Scale to set the vertical scale of the operation results. For setting methods,
refer to the descriptions in "
Parameter Setting Method". You can also enable
the touch screen and use the "Pinch & Stretch" gesture to adjust the vertical
scale.
Press AutoSetting to adjust the vertical scale and offset of the operation
results to an optimal value based on the current configuration, so as to better
observe.
Press More Smooth to set the number of smoothing times in differential
operation. For setting methods, refer to descriptions in "
Parameter Setting
Method".
Note: As the high-frequency component of waveform data has a great impact
on the differential operation, you need to adjust the parameter "Smooth" to a
larger value, so as to restore the low-frequency component.
Press More Invert to enable or disable the inverted display function of the
waveform.
Press More Label to set the label for math operation results. For details,
refer to descriptions in "
Math Operation Label".
Press More Color Grade to enable or disable the color grade display of the
operation results. When enabled, different colors are displayed on the screen to
indicate the times of data acquisition or acquisition probability.
Press More Reset ColorGrade to clear the color grade display and display
the color grade again.
Sqrt
Calculates the square root of the selected source point by point and displays the
results.
Press Math Math1 Operator to select "Sqrt":
Press Operation continuously to enable or disable the "Sqrt" operation
function.
Press SourceA, and rotate the multifunction knob to select the channel for
Source A. Then press down the knob to select the channel. You can also press
SourceA continuously or enable the touch screen to select the channel. The
Tip
Because differentiation is very sensitive to noise, it is helpful to set acquisition
mode to "Average".