Instructions
Table Of Contents
- Guaranty and Declaration
- Safety Requirement
- MSO5000-E Series Overview
- Document Overview
- Quick Start
- General Inspection
- Appearance and Dimensions
- To Prepare for Use
- Front Panel Overview
- Rear Panel Overview
- Front Panel Function Overview
- User Interface
- Touch Screen Controls
- Parameter Setting Method
- To Use the Kensington Security Lock
- To Use the Built-in Help System
- To View the Option Information and the Option Installation
- To Set the Vertical System
- To Set the Horizontal System
- To Set the Sample System
- To Trigger the Oscilloscope
- Trigger Source
- Trigger LEVEL/Threshold Level
- Trigger Mode
- Trigger Coupling
- Trigger Holdoff
- Noise Rejection
- Trigger Type
- Edge Trigger
- Pulse Trigger
- Slope Trigger
- Video Trigger
- Pattern Trigger
- Duration Trigger
- Timeout Trigger
- Runt Trigger
- Window Trigger
- Delay Trigger
- Setup/Hold Trigger
- Nth Edge Trigger
- RS232 Trigger (Option)
- I2C Trigger (Option)
- SPI Trigger (Option)
- CAN Trigger (Option)
- FlexRay Trigger (Option)
- LIN Trigger (Option)
- I2S Trigger (Option)
- MIL-STD-1553 Trigger (Option)
- Zone Trigger
- Trigger Output Connector
- Operations and Measurements
- Digital Voltmeter (DVM) and Frequency Counter
- Power Analysis (Option)
- Histogram Analysis
- Digital Channel
- Protocol Decoding
- Reference Waveform
- To Enable Ref Function
- To Select the Reference Channel
- To Select the Ref Source
- To Adjust the Ref Waveform Display
- To Save to Internal Memory
- To Clear the Display of the Reference Waveform
- To View Details of the Reference Waveform
- To Reset the Reference Waveform
- Color Setting
- Label Setting
- To Export to Internal or External Memory
- To Import from Internal or External Memory
- Pass/Fail Test
- Waveform Recording & Playing
- Search and Navigation Function
- Display Control
- Function/Arbitrary Waveform Generator (Option)
- Store and Load
- System Utility Function Setting
- Remote Control
- Troubleshooting
- Appendix
- Index

RIGOL Chapter 5 To Trigger the Oscilloscope
5-40 MSO5000-E User Guide
Trigger Condition:
Press When to select the desired trigger condition.
SOF: triggers at the start of a frame.
EOF: triggers at the end of a frame.
Remote ID: triggers on the remote frame with the specified ID. Set the following
parameters:
Press Extended ID to enable or disable the extended ID.
Press Bit X to set the data bit that needs to be operated on. For setting
methods, refer to descriptions in "
I2C Trigger (Option)".
Overload: triggers on the CAN overload frames.
Frame ID: triggers on the data frames with the specified ID. Set the following
parameters:
Press Extended ID to enable or disable the extended ID.
Press Bit X to set the data bit that needs to be operated on. For setting
methods, refer to descriptions in "
I2C Trigger (Option)".
Frame Data: triggers on the data frames with specified data. Set the following
parameters:
Press Bit X to set the data bit that needs to be operated on. For setting
methods, refer to descriptions in "
I2C Trigger (Option)".
Press Bytes, then rotate the multifunction knob or use the numeric
keypad to set the length of the data. Its range is from 1 to 8.
Data&ID: triggers on the data frames with the specified ID and data. Set the
following parameters:
Press Define to select "Data" or "ID".
When you select "Data", set Bit X and Bytes. For detailed settings, refer to
parameter settings for "Frame Data".
When you select "ID", set Extended ID and Bit X. For detailed settings,
refer to parameter settings for "Frame ID".
Frame Error: triggers on the error frame.
Bit Fill: triggers on the error frame with the bit fill.
Answer Error: triggers on the answer error frame.
Check Error: triggers on the check error frame.
Format Error: triggers on the format error frame.
Random Error: triggers on random error frame, such as the format error frame,
answer error frame, etc.
Signal Type:
Press More Signal Type to select the desired signal type.
CAN_H: indicates the actual CAN_H bus signal.
CAN_L: indicates the actual CAN_L bus signal.
TX/RX: indicates the Transmit signal and Receive signal from the CAN bus
transceiver.
DIFF: The CAN differential bus signals connected to an analog source channel by
using a differential probe. Connect the probe's positive lead to the CAN_H bus
signal and connect the negative lead to the CAN_L bus signal.