Instructions
Table Of Contents
- Guaranty and Declaration
- Safety Requirement
- MSO5000-E Series Overview
- Document Overview
- Quick Start
- General Inspection
- Appearance and Dimensions
- To Prepare for Use
- Front Panel Overview
- Rear Panel Overview
- Front Panel Function Overview
- User Interface
- Touch Screen Controls
- Parameter Setting Method
- To Use the Kensington Security Lock
- To Use the Built-in Help System
- To View the Option Information and the Option Installation
- To Set the Vertical System
- To Set the Horizontal System
- To Set the Sample System
- To Trigger the Oscilloscope
- Trigger Source
- Trigger LEVEL/Threshold Level
- Trigger Mode
- Trigger Coupling
- Trigger Holdoff
- Noise Rejection
- Trigger Type
- Edge Trigger
- Pulse Trigger
- Slope Trigger
- Video Trigger
- Pattern Trigger
- Duration Trigger
- Timeout Trigger
- Runt Trigger
- Window Trigger
- Delay Trigger
- Setup/Hold Trigger
- Nth Edge Trigger
- RS232 Trigger (Option)
- I2C Trigger (Option)
- SPI Trigger (Option)
- CAN Trigger (Option)
- FlexRay Trigger (Option)
- LIN Trigger (Option)
- I2S Trigger (Option)
- MIL-STD-1553 Trigger (Option)
- Zone Trigger
- Trigger Output Connector
- Operations and Measurements
- Digital Voltmeter (DVM) and Frequency Counter
- Power Analysis (Option)
- Histogram Analysis
- Digital Channel
- Protocol Decoding
- Reference Waveform
- To Enable Ref Function
- To Select the Reference Channel
- To Select the Ref Source
- To Adjust the Ref Waveform Display
- To Save to Internal Memory
- To Clear the Display of the Reference Waveform
- To View Details of the Reference Waveform
- To Reset the Reference Waveform
- Color Setting
- Label Setting
- To Export to Internal or External Memory
- To Import from Internal or External Memory
- Pass/Fail Test
- Waveform Recording & Playing
- Search and Navigation Function
- Display Control
- Function/Arbitrary Waveform Generator (Option)
- Store and Load
- System Utility Function Setting
- Remote Control
- Troubleshooting
- Appendix
- Index

RIGOL Chapter 5 To Trigger the Oscilloscope
5-26 MSO5000-E User Guide
Edge Type:
Press Slope, and then rotate the multifunction knob to select the input signal
the edge on which the oscilloscope triggers. You can also press the Slope key
continuously or enable the touch screen to select the desired edge type.
Rising: triggers on the rising edge of the input signal when the voltage level
is higher than the preset high trigger level.
Falling: triggers on the falling edge of the input signal when the voltage level
is lower than the preset low trigger level.
Either: triggers on the rising or falling edge of the input signal when the
voltage level meets the preset trigger level.
Trigger Position:
After selecting the window type, press Position and then rotate the multifunction
knob
to select the trigger position so as to further specify the time point of
trigger. Press down the knob to select the trigger position. You can also press
Position continuously or enable the touch screen to select it.
Enter: triggers when the input signal enters the specified trigger level range.
Exit: triggers when the input signal exits the specified trigger level range.
Time: triggers when the accumulated hold time since the input signal entered
the specified trigger level range is equal to the window time. When this type is
selected, press Time and rotate the multifunction knob to set the window
time. You can also use the numeric keypad to set it. Its available range is from 8
ns to 10 s.
Level Selection and Adjustment of Trigger Level:
After the trigger condition setting is completed, you need to adjust the trigger level
to correctly trigger the signal and obtain a stable waveform. Press Level Select,
then rotate the multifunction knob
to select the desired level type for adjustment,
and then press down the knob to select the type. You can also press Level Select
continuously to select the level type or enable the touch screen to tap the desired
level type and select it.
Level A: only adjusts the upper limit of the trigger level, and the lower limit of
the trigger level remains unchanged.
Level B: only adjusts the lower limit of the trigger level, and the upper limit of
the trigger level remains unchanged.
Level AB: adjusts the upper and lower limits of the trigger level at the same time,
and the trigger level deviation (the difference between the upper limit and lower
limit of the trigger level) remains unchanged.
Note: In Window trigger, pressing Trigger
LEVEL can quickly switch the
current level adjustment type.
Rotate the Trigger LEVEL knob to adjust the corresponding trigger level.
During the adjustment of trigger level, two trigger level lines appear on the screen,
and they move up and down with the change of the trigger level. At the same time,
the real-time trigger level and slope information is displayed at the lower-left corner