Instructions
Table Of Contents
- Guaranty and Declaration
- Safety Requirement
- MSO5000-E Series Overview
- Document Overview
- Quick Start
- General Inspection
- Appearance and Dimensions
- To Prepare for Use
- Front Panel Overview
- Rear Panel Overview
- Front Panel Function Overview
- User Interface
- Touch Screen Controls
- Parameter Setting Method
- To Use the Kensington Security Lock
- To Use the Built-in Help System
- To View the Option Information and the Option Installation
- To Set the Vertical System
- To Set the Horizontal System
- To Set the Sample System
- To Trigger the Oscilloscope
- Trigger Source
- Trigger LEVEL/Threshold Level
- Trigger Mode
- Trigger Coupling
- Trigger Holdoff
- Noise Rejection
- Trigger Type
- Edge Trigger
- Pulse Trigger
- Slope Trigger
- Video Trigger
- Pattern Trigger
- Duration Trigger
- Timeout Trigger
- Runt Trigger
- Window Trigger
- Delay Trigger
- Setup/Hold Trigger
- Nth Edge Trigger
- RS232 Trigger (Option)
- I2C Trigger (Option)
- SPI Trigger (Option)
- CAN Trigger (Option)
- FlexRay Trigger (Option)
- LIN Trigger (Option)
- I2S Trigger (Option)
- MIL-STD-1553 Trigger (Option)
- Zone Trigger
- Trigger Output Connector
- Operations and Measurements
- Digital Voltmeter (DVM) and Frequency Counter
- Power Analysis (Option)
- Histogram Analysis
- Digital Channel
- Protocol Decoding
- Reference Waveform
- To Enable Ref Function
- To Select the Reference Channel
- To Select the Ref Source
- To Adjust the Ref Waveform Display
- To Save to Internal Memory
- To Clear the Display of the Reference Waveform
- To View Details of the Reference Waveform
- To Reset the Reference Waveform
- Color Setting
- Label Setting
- To Export to Internal or External Memory
- To Import from Internal or External Memory
- Pass/Fail Test
- Waveform Recording & Playing
- Search and Navigation Function
- Display Control
- Function/Arbitrary Waveform Generator (Option)
- Store and Load
- System Utility Function Setting
- Remote Control
- Troubleshooting
- Appendix
- Index

RIGOL Chapter 5 To Trigger the Oscilloscope
5-24 MSO5000-E User Guide
descriptions in "Trigger Source". The current trigger source is displayed at the
upper-right corner of the screen.
Note: Only when we select the channel (that has been input with signals) as the
trigger source, can we obtain a stable trigger.
Polarity:
Press Polarity continuously to select the pulse polarity of Runt trigger.
Positive: triggers on the positive runt pulse.
Negative: triggers on the negative runt pulse.
Trigger Condition:
Press When and rotate the multifunction knob to set the trigger conditions of
Runt trigger. You can also press When continuously or enable the touch screen to
select it.
None: indicates not setting the trigger condition of Runt trigger.
>: triggers when the runt pulse width is greater than the lower limit of pulse
width. Press Lower to set the minimum pulse width of Runt trigger.
<: triggers when the runt pulse width is smaller than the upper limit of pulse
width. Press Upper to set the maximum pulse width of Runt trigger.
<>: triggers when the runt pulse width is greater than the lower limit and
smaller than the upper limit of pulse width. Press Upper to set the maximum
pulse width of Runt trigger. Press Lower to set the minimum pulse width of
Runt trigger.
Note: The lower limit of the pulse width must be smaller than the upper limit.
Level Selection and Adjustment of Trigger Level:
After the trigger condition setting is completed, you need to adjust the trigger level
to correctly trigger the signal and obtain a stable waveform. Press Level Select
(when "<>" is selected as the trigger condition, Level Select is a sub-menu under
More), then rotate the multifunction knob to select the desired level type for
adjustment, and then press down the knob to select the type. You can also press
Level Select continuously to select the level type or enable the touch screen to tap
the desired level type and select it.
Level A: only adjusts the upper limit of the trigger level, and the lower limit of the
trigger level remains unchanged.
Level B: only adjusts the lower limit of the trigger level, and the upper limit of the
trigger level remains unchanged.
Level AB: adjusts the upper and lower limits of the trigger level at the same time, and
the trigger level deviation (the difference between the upper limit and lower limit of
the trigger level) remains unchanged.
Note: In Runt trigger, pressing Trigger
LEVEL can quickly switch the current
level adjustment type.
Rotate the Trigger LEVEL knob to adjust the corresponding trigger level.
During the adjustment of trigger level, two trigger level lines appear on the screen,