Instructions
Chapter 2 Command System  RIGOL 
MSO5000-E Programming Guide  2-181  
:TRIGger:WINDows 
Command List: 
  :TRIGger:WINDows:SOURce 
  :TRIGger:WINDows:SLOPe 
  :TRIGger:WINDows:POSition 
  :TRIGger:WINDows:TIME 
  :TRIGger:WINDows:ALEVel 
  :TRIGger:WINDows:BLEVel 
:TRIGger:WINDows:SOURce 
Syntax  :TRIGger:WINDows:SOURce <source> 
:TRIGger:WINDows:SOURce? 
Description 
Sets or queries the trigger source of Window trigger. 
Parameter 
Name 
Type 
Range 
Default 
<source> 
Discrete 
{CHANnel1|CHANnel2} 
CHANnel1 
Return 
Format 
The query returns CHAN1 or CHAN2. 
Example 
:TRIGger:WINDows:SOURce CHANnel2       /*Sets the trigger source to CHANnel2.*/ 
:TRIGger:WINDows:SOURce?  /*The query returns CHAN2.*/ 
:TRIGger:WINDows:SLOPe 
Syntax  :TRIGger:WINDows:SLOPe <type> 
:TRIGger:WINDows:SLOPe? 
Description 
Sets or queries the window type of Window trigger. 
Parameter 
Name 
Type 
Range 
Default 
<type> 
Discrete 
{POSitive|NEGative|RFALl} 
POSitive 
Remarks    POSitive: triggers on the rising edge of the input signal when the voltage level is 
greater than the preset high trigger level. 
  NEGative: triggers on the falling edge of the input signal when the voltage level 
is smaller than the preset low trigger level. 
  RFALl: triggers on either the rising or falling edge of the input signal when the 
voltage level meets the preset trigger level. 
Return 
Format 
The query returns POS, NEG, or RFAL. 
Example 
:TRIGger:WINDows:SLOPe NEGative    /*Sets the window type to NeGative.*/ 
:TRIGger:WINDows:SLOPe?  /*The query returns NEG.*/ 