Instructions

Chapter 2 Command System RIGOL
MSO5000-E Programming Guide 2-169
:TRIGger:VIDeo:MODE
Syntax :TRIGger:VIDeo:MODE <mode>
:TRIGger:VIDeo:MODE?
Description
Sets or queries the sync type of Video trigger.
Parameter
Name
Type
Range
Default
<mode>
Discrete
{ODDField|EVENfield|LINE|ALINes}
ALINes
Remarks ODDField: indicates the odd field. The oscilloscope triggers on the rising edge of
the first ramp waveform in the odd field. It is available when the video standard is
NTSC or PAL/SECAM.
EVENfield: indicates the even field. The oscilloscope triggers on the rising edge of
the first ramp waveform in the even field. It is available when the video standard is
NTSC or PAL/SECAM.
LINE: for NTSC and PAL/SECAM video standards, the oscilloscope triggers on the
specified line in the odd or even field. For 480P and 576P video standards, the
oscilloscope triggers on the specified line.
ALINes: triggers on all the horizontal sync pulses.
Return
Format
The query returns ODDF, EVEN, LINE, or ALIN.
Example
:TRIGger:VIDeo:MODE ODDField /*Sets the sync type to ODDField.*/
:TRIGger:VIDeo:MODE? /*The query returns ODDF.*/
:TRIGger:VIDeo:LINE
Syntax :TRIGger:VIDeo:LINE <line>
:TRIGger:VIDeo:LINE?
Description
Sets or queries the sync type of Video trigger to Line.
Parameter
Name
Type
Range
Default
<line>
Integer
Refer to Remarks
1
Remarks PALSecam: 1 to 625
NTSC: 1 to 525
480P: 1 to 525
576P: 1 to 625
720P60: 1 to 750
720P50: 1 to 750
720P30: 1 to 750
720P251 to 750
720P24: 1 to 750
1080P60: 1 to 1125
1080P50: 1 to 1125
1080P30: 1 to 1125
1080P25: 1 to 1125