Instructions

Chapter 2 Command System RIGOL
MSO5000-E Programming Guide 2-151
:SYSTem:OPTion:UNINSTall
Syntax
:SYSTem:OPTion:UNINSTall
Description
Uninstalls all the official options.
Remarks
After the option has been uninstalled, you need to restart the instrument.
:SYSTem:OPTion:STATus?
Syntax
:SYSTem:OPTion:STATus? <type>
Description
Queries the status of the options of the oscilloscope.
Parameter
Name
Type
Range
Default
<type> Discrete
{RL1|BND|COMP|EMBD|AUTO|FLEX|AUDI
O|AERO|AWG|PWR}
——
说明
RL1: memory depth option. Maximum memory depth up to 100 Mpts.
BND: function and application bundle option, MSO5000-COMP, MSO5000-EMBD,
MSO5000-AUTO, MSO5000-FLEX, MSO5000-AUDIO, MSO5000-AERO,
MSO5000-E-AWG, and MSO5000-PWR.
COMP: PC serial bus trigger and analysis option (RS232/UART).
EMBD: embedded serial bus trigger and analysis option (I2C and SPI).
AUTO: auto serial bus trigger and analysis option (CAN and LIN).
FLEX: FlexRay serial bus trigger and analysis option (FlexRay).
AUDIO: audio serial bus trigger and analysis (I2S).
AERO: MIL-STD-1553 serial bus trigger and analysis option (MIL-STD-1553).
AWG: 25 MHz Arbitrary Waveform Generator.
PWR: built-in power analysis option.
Return
Format
The query returns 0 or 1.
0: indicates that the option is not installed.
1: indicates that the official option has been installed.
:SYSTem:RAMount?
Syntax
:SYSTem:RAMount?
Description
Queries the number of analog channels of the current instrument.
Return
Format
The query returns the number of analog channels of the current instrument in integer.