Instructions
Chapter 2 Command System  RIGOL 
MSO5000-E Programming Guide  2-151  
:SYSTem:OPTion:UNINSTall 
Syntax 
:SYSTem:OPTion:UNINSTall 
Description 
Uninstalls all the official options. 
Remarks 
After the option has been uninstalled, you need to restart the instrument. 
:SYSTem:OPTion:STATus? 
Syntax 
:SYSTem:OPTion:STATus? <type> 
Description 
Queries the status of the options of the oscilloscope. 
Parameter 
Name 
Type 
Range 
Default 
<type>  Discrete 
{RL1|BND|COMP|EMBD|AUTO|FLEX|AUDI
O|AERO|AWG|PWR} 
—— 
说明
  RL1: memory depth option. Maximum memory depth up to 100 Mpts. 
  BND: function and application bundle option, MSO5000-COMP, MSO5000-EMBD, 
MSO5000-AUTO, MSO5000-FLEX, MSO5000-AUDIO, MSO5000-AERO, 
MSO5000-E-AWG, and MSO5000-PWR. 
  COMP: PC serial bus trigger and analysis option (RS232/UART). 
  EMBD: embedded serial bus trigger and analysis option (I2C and SPI). 
  AUTO: auto serial bus trigger and analysis option (CAN and LIN). 
  FLEX: FlexRay serial bus trigger and analysis option (FlexRay). 
  AUDIO: audio serial bus trigger and analysis (I2S). 
  AERO: MIL-STD-1553 serial bus trigger and analysis option (MIL-STD-1553). 
  AWG: 25 MHz Arbitrary Waveform Generator. 
  PWR: built-in power analysis option. 
Return 
Format 
The query returns 0 or 1. 
0: indicates that the option is not installed. 
1: indicates that the official option has been installed.
:SYSTem:RAMount? 
Syntax 
:SYSTem:RAMount? 
Description 
Queries the number of analog channels of the current instrument. 
Return 
Format 
The query returns the number of analog channels of the current instrument in integer.  