Chapter 2 Command System RIGOL
MSO5000-E Programming Guide 2-149
KFUNction: press down the multifunction knob.
TLEVel: press down the trigger level adjustment knob, i.g. set the trigger level to
50% of the peak-peak value.
TMENu: press the MENU key in the trigger control area.
TMODe: press the MODE key in the trigger control area.
RSTop: press the RUN/STOP key.
Example
:SYSTem:KEY:PRESs <CH1>/*Press the CH1 key on the front panel of the oscilloscope.*/
:SYSTem:KEY:INCRease
:SYSTem:KEY:DECRease
Syntax :SYSTem:KEY:INCRease <key>[,<val>]
:SYSTem:KEY:DECRease <key>[,<val>]
Description Rotate the specified knob clockwise.
Rotate the specified knob counterclockwise.
Parameter
Name
Type
Range
Default
<key> Discrete
{VOFFset1|VOFFset2|VSCale1|VSCale2|HSCal
e|HPOSition|KFUNction|TLEVel}
——
<val> Integer
Determined by the range of the parameter
<key> and the current setting of the
parameter.
1
Remarks The value of the parameter <key> corresponds to the specified knob on the front
panel. The definitions for the knob are as follows:
VOFFset1: the CH1 vertical offset knob.
VOFFset2: the CH2 vertical offset knob.
VSCale1: the vertical scale adjustment knob of CH1.
VSCale2: the vertical scale adjustment knob of CH2.
HSCale: the horizontal time base adjustment knob.
HPOSition: the horizontal position adjustment knob.
KFUNction: the multifunction knob.
TLEVel: the trigger level adjustment knob.
The parameter <val> defines the specified rotation times of the knob. By default,
it is 1.
Example :SYSTem:KEY:INCRease VOFFset2 /*Rotates the vertical offset adjustment knob of
CH2 clockwise.*/
:SYSTem:LANGuage
Syntax :SYSTem:LANGuage <language>
:SYSTem:LANGuage?
Description
Sets or queries the system language.
Parameter
Name
Type
Range
Default
<language> Discrete
{SCHinese|TCHinese|KORean|JAPanese|EN
GLish|GERMan|PORTuguese|POLish|FRENch
|RUSSian|SPAN|THAI|INDonesian }
SCHinese