Instructions
RIGOL Chapter 2 Performance Verification Test Overview
2-4 MSO5000-E Performance Verification Guide
current vertical scale respectively.
3) Repeat Step 3-7 and record the test results.
4) Calculate the relative error of each vertical scale: |(Vavg1 - Vavg2) - (Vout1 -
Vout2)|/Full Scale × 100%.
10. Turn off CH1. Test the relative error of each scale of CH2 using the method above and record the
test results.
Test Record Form
Channel
Vertical
Scale
Test Result
Limit Pass/Fail
Vavg1 Vavg2
Calculation
Result
[1]
CH1
500 uV/div
≤3%
1 mV/div
2 mV/div
5 mV/div
10 mV/div
20 mV/div
50 mV/div
100 mV/div
200 mV/div
500 mV/div
1 V/div
2 V/div
5 V/div
10 V/div
CH2
500 uV/div
1 mV/div
2 mV/div
5 mV/div
10 mV/div
20 mV/div
50 mV/div
100 mV/div
200 mV/div
500 mV/div
1 V/div
2 V/div
5 V/div
10 V/div
Note
[1]
: The calculation formula is |(Vavg1 - Vavg2) - (Vout1 - Vout2)|/Full Scale × 100%; wherein, Vout1 and Vout2
are 3 × the current vertical scale and -3 × the current vertical scale respectively.