Instructions
RIGOL Chapter 2 Performance Verification Test Overview
2-2 MSO5000-E Performance Verification Guide
Impedance Test
Specification
Input Impedance
Analog Channel 1 MΩ: 0.99 MΩ to 1.01 MΩ
Test Connection Diagram
Figure 2-1 Impedance Test Connection Diagram
Test Procedures
1. Connect the active head of Fluke 9500B to CH1 of the oscilloscope, as shown in the figure above.
2. Configure the oscilloscope:
1) Press 1 in the vertical control area (Vertical) on the front panel to enable CH1.
2) Press 1 Impedance to set the input impedance of CH1 to 1 MΩ.
3) Set the vertical scale of CH1 to 100 mV/div.
3. Turn on Fluke 9500B; set its impedance to 1 MΩ and select the resistance measurement function.
Read and record the resistance measured.
4. Adjust the vertical scale of CH1 of the oscilloscope to 500 mV/div; read and record the resistance
measured.
5. Turn off CH1. Measure the resistance of CH2 using the method above and record the
measurement results.
Test Record Form
Channel
Vertical Scale
Test Result
Limit
Pass/Fail
CH1
100 mV/div
0.99 MΩ to 1.01 MΩ
500 mV/div
CH2
100 mV/div
500 mV/div
Fluke 9500B
MSO5000-E