Instructions
Chapter 2 Performance Verification Test Overview RIGOL
MSO5000-E Performance Verification Guide
2-1
Chapter 2 Performance Verification Test
This chapter takes MSO5152-E as an example to illustrate the performance verification test methods
and procedures of MSO5000-E series digital oscilloscope. Fluke 9500B is used in this manual for the
tests. You can also use other devices that fulfill the "Specification" in Table 2-1.
Table 2-1 Test Devices Required
Device Specification
Recommended
Model
Oscilloscope
Calibrator
DC output voltage range:
1 MΩ: 1 mV to 200 V
50 Ω: 1 mV to 5 V
Fast edge signal rise time: ≤ 150 ps
Fluke 9500B
Note:
1. Make sure that the oscilloscope passes the self-test and self-calibration before executing the
performance verification tests.
2. Make sure that the oscilloscope has been warmed up for at least 30 minutes before executing any
of the following tests.
3. Please reset the instrument to the factory setting before or after executing any of the following
tests.