Instructions
RIGOL  Chapter 5 To Trigger the Oscilloscope 
5-6  MSO5000 User's Guide 
  AC: blocks the DC components and attenuates signals. 
  LFR: blocks the DC components and rejects the low-frequency components. 
  HFR: rejects the high frequency components. 
Trigger Holdoff 
Trigger holdoff can be used to stably trigger on complex repetitive waveforms that 
have multiple edges (or other events) between waveform repetitions (such as pulse 
series). Holdoff time indicates the time that the oscilloscope waits for re-arming the 
trigger module after generating a correct trigger. The oscilloscope will not trigger 
even if the trigger condition is met during the holdoff time and will only re-arm the 
trigger module after the holdoff time expires. 
For example, to stably trigger the repetitive pulse series as shown in the figure below, 
the holdoff time should be set to a value that is greater than t1 and smaller than t2. 
Figure 5-2 Schematic Diagram of Trigger Holdoff 
In the trigger control area (Trigger) on the front panel, press Menu  Holdoff (for 
some triggers, Holdoff is a sub menu under More, at this time, press Menu  
More  Holdoff) and then rotate the multifunction knob    at a small step to 
modify the holdoff time (the holdoff to this time when the waveforms are stably 
triggered; by default, the holdoff time is 16 ns). Also, you can use the numeric 
keypad to set the holdoff time. The adjustable range of holdoff time is from 8 ns to 
10 s.   
Tip 
When "AC" or "LFR" is selected under Coupling, no trigger level lines and trigger 
icons are displayed. When you adjust the trigger level, you can only see the changes 
of the trigger level values at the upper-right corner of the screen. 
Holdoff Time 
Trigger Position 
t1 
t2 










