Instructions
Chapter 5 To Trigger the Oscilloscope    RIGOL 
MSO5000 User's Guide  5-5 
are forced and acquisitions are made so as to display the waveforms. This trigger 
mode should be used when the signal level is unknown or the DC should be 
displayed as well as when forcible trigger is not necessary as the trigger condition 
always occurs. 
  Normal: In this trigger mode, triggers and acquisitions only occur when the specified 
trigger conditions are found. This trigger mode should be used when the signal is 
with low repetition rate or only the event specified by the trigger setting needs to be 
sampled as well as when auto trigger should be prevented to acquire stable display. 
  Single:  In  this  trigger  mode,  the  oscilloscope  performs  a  single  trigger  and 
acquisition when the specified trigger conditions are found, and then stops. This 
trigger mode should be used when you need to perform a single acquisition of 
the specified event and analyze the acquisition result (you can pan and zoom 
the currently displayed waveform, and the subsequent waveform data will not 
overwrite  the current waveform).  After  a  single  trigger mode is initiated, the 
operating status of the oscilloscope is in "STOP" state.   
Note:  In  "Normal"  and  "Single"  trigger  modes,  pressing  Force  can  generate  a 
trigger signal forcibly. 
In addition, you can enable the touch  screen to set the trigger mode. Enable the 
touch  screen  and  tap  the  trigger  setting  label    at  the  upper-right 
corner of the screen. Then the following window is displayed. Tap to select "Auto", 
"Normal", or "Single" as the trigger mode. 
Trigger Coupling 
Trigger coupling decides which kind of components will be transmitted to the trigger 
module. Please distinguish it from "Channel Coupling". 
Note: Trigger coupling is only valid when the trigger type is Edge and the trigger 
source is an analog channel. 
Press Menu  Coupling in the trigger control area (Trigger) on the front panel. 
Then, rotate the multifunction knob    to select the desired coupling mode (by 
default, it is DC). You can also press Coupling continuously to select it or enable the 
touch screen to select the desired coupling mode with touch gestures. 
  DC: allows DC and AC components to pass the trigger path. 