RIGOL  Chapter 2 To Set the Vertical System 
2-8  MSO5000 User's Guide 
(a) "Invert" Off 
(b) "Invert" On 
Figure 2-1 Waveform Invert 
Note: When the waveform invert is enabled, the trigger edge or the trigger polarity 
will change (e.g. Edge trigger, Pulse trigger, or Slope trigger). 
Amplitude Unit 
Select the amplitude display unit for the current channel. The available units are W, A, 
V, and U. When the unit is changed, the unit related to the relevant functions of the 
channel will change accordingly. 
Press 1  More  Unit, press Unit continuously or rotate the multifunction knob 
  to select a unit. You can also enable the touch screen to select the desired unit 
with touch gestures. The default unit for the value is V. 
Channel Delay 
When using an oscilloscope for actual measurement, the transmission delay of the 
probe cable may bring relatively greater errors (zero offset). MSO5000 allows you to 
set a delay time for calibrating the zero offset of the corresponding channel. Zero 
offset is defined as the offset of the crossing point of the waveform and trigger level 
line relative to the trigger position, as shown in Figure 2-2. 