Instructions
Chapter 6 Operations and Measurements    RIGOL 
MSO5000 User's Guide  6-29 
Measurement Parameter 
Press Measure on the front panel to enter the measurement setting menu. You can 
set the measurement source, enable or disable the all measurement function, the 
statistical analysis function, and etc. You can also quickly make measurements for 41 
waveform parameters. The measurement results will be displayed on the screen. The 
measurement results on the screen are always marked in the same color as the 
current measurement channel. You can also enable the touch screen to tap the auto 
measurement label "Measure" at the top of the screen to enter the measurement 
setting menu. 
Note: If there is no signal input for the current source or the measurement result is 
not within the valid range (too large or too small), then the measurement results are 
invalid, and "*****" is displayed on the screen. Please re-input the signal or set the 
signal. 
Time Parameters 
Figure 6-1 Time Parameters 
1. Period:  defined  as  the  time  between  the  middle  threshold  points  of  two 
consecutive, like-polarity edges. 
2. Frequency: defined as the reciprocal of period. 
3. Rise Time: indicates the time for the signal amplitude to rise from the threshold 
lower limit to the threshold upper limit. 
4. Fall Time: indicates the time for the signal amplitude to rise from the threshold 
upper limit to the threshold lower limit. 
5. +Width: indicates the time between the threshold middle value of a rising edge 
to the threshold middle value of the next falling edge. 
Fall Time 
Threshold 
Upper Limit 
Threshold 
Middle Value 
Threshold 
Lower Limit 
+Width 
-Width 
Period 
Rise Time 