RIGOL  Chapter 5 To Trigger the Oscilloscope 
5-22  MSO5000 User's Guide 
Runt Trigger 
This trigger mode is used to trigger pulses that pass through one trigger level but fail 
to pass through another trigger level, as shown in the figure below. 
Figure 5-9 Runt Trigger 
Trigger Type: 
Press Type, and then rotate the multifunction knob    to select "Runt". Press down 
the knob to select the trigger type. Then, the current trigger setting information is 
displayed at the upper-right corner of the screen, as shown in the figure below. You 
can also press Type continuously to select the trigger type or enable the touch 
screen to tap the desired trigger type and select it. 
Source Selection: 
Press Source to switch the trigger source and select CH1-CH4. For details, refer to 
descriptions in  "Trigger  Source". The current  trigger  source  is displayed  at  the 
upper-right corner of the screen. 
Note: Only when we select the channel (that has been input with signals) as the 
trigger source, can we obtain a stable trigger. 
Polarity: 
Press Polarity continuously to select the pulse polarity of Runt trigger. 
  Positive: triggers on the positive runt pulse. 
  Negative: triggers on the negative runt pulse. 
Trigger Condition: 
Press When and rotate the multifunction knob    to set the trigger conditions of 
Runt trigger. You can also press When continuously or enable the touch screen to 
select it. 
  None: indicates not setting the trigger condition of Runt trigger. 
  >: triggers when the runt pulse width is greater than the lower limit of pulse 
width. Press Lower to set the minimum pulse width of Runt trigger.   
  <: triggers when the runt pulse width is smaller than the upper limit of pulse 
width. Press Upper to set the maximum pulse width of Runt trigger.   
Lower Limit of 
the Trigger Level 
Upper Limit of the 
Trigger Level 
Negative Runt Pulse 
Positive Runt Pulse 